datasheet,schematic,electronic components, service manual,repairs,tv,monitor,service menu,pcb design
Schematics 4 Free
Service manuals, schematics, documentation, programs, electronics, hobby ....


registersend pass
Bulgarian - schematics repairs service manuals SearchBrowseUploadWanted

Now downloading free:Agilent ARINC 429 mask files and Application Note ARINC 429 Mask Files and Application Note

Agilent ARINC 429 mask files and Application Note ARINC 429 Mask Files and Application Note free download

Various electronics service manuals

File information:
File name:ARINC_429_mask_files_and_Application_Note ARINC 429 Mask Files and Application Note.zip
[preview ARINC 429 mask files and Application Note ARINC 429 Mask Files and Application Note]
Size:2229 kB
Extension:zip
Mfg:Agilent
Model:ARINC 429 mask files and Application Note ARINC 429 Mask Files and Application Note 🔎
Original:ARINC 429 mask files and Application Note ARINC 429 Mask Files and Application Note 🔎
Descr: Agilent ARINC_429_mask_files_and_Application_Note ARINC 429 Mask Files and Application Note.zip
Group:Electronics > Other
Uploaded:04-08-2022
User:Anonymous
Multipart:No multipart

Information about the files in archive:
Decompress result:OK
Extracted files:1
File name ARINC_429_mask_files_and_Application_Note ARINC 429 Mask Files and Application Note.zip

ARINC 429 Eye-diagram and Pulse-shape Mask Testing Application Note Probing the Differential ARINC 429 Bus ARINC 429 mask testing is based on capturing and overlaying all "1" and/or "0" bits of the differential bus. The differential bus must be probed using a differential active probe. Agilent recommends using the N2791A 25-MHz differential active probe shown in Figure 1. The attenuation setting on the probe should be set to 10:1 (not 100:1). The output of the probe should be terminated into the scope's default 1-M input termination.. Introduction Eye-diagram mask testing is used in a broad range of today's serial bus applications. An eye-diagram is basically an overlay of all bits captured by the scope to show when bits are valid and not valid. This provides a composite picture of the overall quality of a system's physical layer characteristics, which includes amplitude variations possibly due to transmission line affects, reflections, system noise, over-shoot, ringing, signal edge timing, and jitter. Eye-diagram and pulse-shape pass/fail mask testing can be performed on differential ARINC 429 signals using an Agilent 3000 or 4000 X-Series oscilloscope licensed with the DSOX3AERO/DSOX4AERO triggering and decode option (MIL-STD 1553 & ARINC 429), along with the DSOX3MASK/ DSOX4MASK mask test option. Various ARINC 429 mask files can be downloaded from Agilent's website at no charge. Save the appropriate ARINC 429 mask files (based on baud rate) to your personal USB memory stick and then insert the memory stick into the scope's front-panel USB port. The following ARINC 429 mask files are available:



>> Download document << eServiceInfo Context Help



Was this file useful ? Share Your thoughts with the other users.

User ratings and reviews for this file:

DateUserRatingComment

Average rating for this file: 0.00 ( from 0 votes)


Similar Service Manuals :
Agilent ReleaseNotes SW ReleaseNotes SW Desktop A050220121101 - Agilent README.txt README DEE A.05.00.2012.0630 - Agilent nc param - Agilent ReleaseNotes SW ReleaseNotes SW A050320130124 - Agilent ReleaseNotes FW A.05.00.2012.0710.txt ReleaseNotes FW A.05.00.2012.0710 - Agilent Release Note FW A.05.04.2013.0328 ReleaseNotes FW A0504 20130328 - Agilent ReleaseNotes SW ReleaseNotes SW A.05.00.2012.0630 -
 FB -  Links -  Info / Contacts -  Forum -   Last SM download :

script execution: 0.07 s