QA-90
User & Service Manual

QA-90 Electrical Safety Analyzer

P/N 11025 V.3.47

Copyright 2004 by METRON. All rights reserved. METRON:
USA ____ 1345 Monroe NW Grand Rapids, MI 49505 Phone: (+1) 888 863-8766 Fax: (+1) 616 454-3350 E-mail: [email protected] FRANCE________ 30, rue Paul Claudel 91000 Evry, France Phone: (+33) 1 6078 8899 Fax: (+33) 1 6078 6839 E-mail: [email protected] GERMANY NORWAY______ Gundastrasse 29 Vegamot 8 D-63762 Grossostheim N-7048 Trondheim Phone: (+49) 6026 993975 Phone: (+47) 7395 4700 Fax: (+49) 6026 977079 Fax: (+47) 7395 4701 E-mail:[email protected] E-mail: [email protected]

Disclaimer
METRON provides this publication as is without warranty of any kind, either express or implied, including but not limited to the implied warranties of merchantability or fitness for any particular purpose. Further, METRON reserves the right to revise this publication and to make changes from time to time to the content hereof, without obligation to METRON or its local representatives to notify any person of such revision or changes. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.

Limited Warranty
METRON warrants that the QA-90 Electrical Safety Analyzer will substantially conform to published specifications and to the documentation, provided that it is used for the purpose for which it was designed. METRON will, for a period of twelve (12) months from date of purchase, replace or repair any defective analyzer, if the fault is due to a manufacturing defect. In no event will METRON or its local representatives be liable for direct, indirect, special, incidental, or consequential damages arising out of the use of or inability to use the QA-90 Electrical Safety Analyzer, even if advised of the possibility of such damages. METRON or its local representatives are not responsible for any costs, loss of profits, loss of data, or claims by third parties due to use of, or inability to use the QA-90 Electrical Safety Analyzer. Neither METRON nor its local representatives will accept, nor be bound by any other form of guarantee concerning the QA-90 Electrical Safety Analyzer other than this guarantee. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.

ii

Table of Contents
MANUAL REVISION RECORD .........................................................................................................V 1. INTRODUCTION ................................................................................................................... 1-1 1.1 1.2 1.3 2. 2.1 2.2 2.3 2.4 3. 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4. 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 5. 5.1 5.2 5.3 5.4 5.5 QA-90 Features.................................................................................................................. 1-1 Specifications..................................................................................................................... 1-1 General Information........................................................................................................... 1-3 Receipt, Inspection and Return.......................................................................................... 2-1 Setup .................................................................................................................................. 2-2 Power ................................................................................................................................. 2-2 Ansur & PRO-Soft............................................................................................................. 2-2 Control Switches and Connections .................................................................................... 3-1 Key Pad Functions............................................................................................................. 3-3 Function of The Bar Code Reader ..................................................................................... 3-3 Menu and Function Keys................................................................................................... 3-4 Display Menus and Messages............................................................................................ 3-4 Measurements with Several Modules in Manual Mode..................................................... 3-9 Storing Setup Parameters in Flash Memory .................................................................... 3-10 Upgrading the QA-90 Software Program ........................................................................ 3-11 Cleaning the QA-90......................................................................................................... 3-14 Test Lead Calibration ........................................................................................................ 4-1 Connecting an Instrument without Patient Inputs.............................................................. 4-2 Connecting an Instrument with Patient Inputs................................................................... 4-3 Power Cable Test............................................................................................................... 4-5 Current Measurement Test (Dual Lead) ............................................................................ 4-6 Voltage Measurement Test (Dual Lead)............................................................................ 4-8 Resistance Measurement (Dual Lead) ............................................................................. 4-10 Connecting Auxiliary Power Source / Isolating Transformer ......................................... 4-11 Required Test Equipment .................................................................................................. 5-1 Preparation and Adjustments............................................................................................. 5-1 Calibration ......................................................................................................................... 5-4 Other Calibration Procedures .......................................................................................... 5-10 Calibration Constants....................................................................................................... 5-12

INSTALLATION..................................................................................................................... 2-1

OPERATING QA-90............................................................................................................... 3-1

EXAMPLE TEST MEASUREMENTS ................................................................................. 4-1

CONTROL AND CALIBRATION........................................................................................ 5-1

iii

6.

THEORY OF OPERATION .................................................................................................. 6-1 6.1 A.1 A.2 A.3 A.4 A.5 Theory of Operation .......................................................................................................... 6-1 Classification of Equipment...................................................................................................1 Tests on Mains Powered Class 1 & 2 Equipment According To IEC 60601.1/UL 2601.1...4 Tests on Internally Powered Equipment According To IEC 60601.1/UL 2601.1...............13 System Tests Based on IEC 60601.1/UL 2601.1.................................................................15 Tests According To VDE 0751:10-1990/2001....................................................................16

APPENDIX A: IEC 60601.1, UL 2601.1 AND VDE 0751 TESTING................................................1

APPENDIX B: ERROR REPORT FORM, QA-90 .............................................................................1 APPENDIX C: SUGGESTION FORM, QA-90 ..................................................................................1

iv

Manual Revision Record
This record page is for recording revisions to your QA-90 User and Service Manual that have been published by METRON or its authorized representatives. We recommend that only the management or facility representative authorized to process changes and revisions to publications: · · · make the pen changes or insert the revised pages; ensure that obsolete pages are withdrawn and either disposed of immediately, or marked as superseded and placed in a superseded document file, and; enter the information below reflecting that the revisions have been entered.
Date Entered Reason Initial Release Signature of Person Entering Change 0

Rev No

3.13-1 3.47-1

04-26-01 01-15-04

General update General update

v

Symbols Used In This Manual
Caution (Refer to accompanying documents) (See NOTE) Caution, Risk of Electrical Shock (see NOTE) Earth (ground) TERMINAL

vi

1. Introduction
This chapter describes the QA-90 Electrical Safety Tester's features and specifications. 1.1 QA-90 Features The QA-90 represents a new generation of safety testing equipment. It is the only automatic safety analyzer that can test units with different classes of protection in one test run (e.g., cardiac float and body float defibrillators). It is simple to use. All you need do is select the type and class of equipment to test. When you press START, QA-90 executes the tests prescribed to the selected standard. Test results may be printed out immediately, or stored internally in the unit for later use. QA-90 has full remote control, and may be operated from PRO-Soft QA-90 software. PRO-Soft QA-90 enables you to make your own test protocols, store the information on disk, and export formatted data to any other database or equipment management program. Individual test sequences may be compiled to satisfy national and international standards. The following standards may be compiled either fully or in part: IEC 60601.1, UL 2601.1, IEC 60601.1.1, UL 2601.1.1, IEC 60601.2.4, IEC 61010-1, EN 60601-1, VDE 0750 Tl/12-91, BS 5724, CAN/CSA-C22.2 No 601.1-M90, AS 3200.1, NZS 6150:1990, VDE 0751-1:10-1990, VDE 0751-1:10-2001, ÖVE 0751, UL 544, HEI 95, HEI 158 among others. 1.2 Specifications
1. Voltage Measurement Measurements may be obtained in the following ways: · · · · Between leads 1 and 2 (in the power contact). Between lead 1 and ground (in the power contact). Between lead 2 and ground (in the power contact). Between input/output E+ and E- (floating inputs/outputs). 0 - 400V true RMS. 0.1V DC - 100 Hz, 1% of full scale +1 LSD 100 Hz-10 kHz, 2% of scale +1 LSD 4 or multiple (LSD = least significant digit)

Range Resolution Accuracy No. of Tests

1-1

2. Current Consumption The current measurement may be executed in lead no. 1 (live). Range 1 Resolution Accuracy No. of Tests Range 2 Resolution Accuracy No. of Tests 3. Protective Earth The test current is selectable from 25A or 1A, delivered from a transformer with a maximum idle voltage of 6V. The measurement can be performed on ground leads or between E+ and E- (floating inputs/outputs). Range Resolution Accuracy No. of Tests 4. Insulating Resistance The measurement of the insulating resistance may be executed between casing and power unit, or between patient module and power unit. Test voltage: No. of Tests: Range Resolution Accuracy Range Resolution Accuracy 5. Leakage Currents All measurements can be performed with a IEC 601.1 filter (patient equivalent), or without (flat frequency response). The filter can be exchanged with filters covering other standards. All measurements can be performed as true RMS measurements, or AC/DC measurements. The following leakage currents can be measured: Ground leakage current Enclosure leakage current Differential current * Substitute equipment lekage current Direct leakage current Leakage current fig. 9 No. of Tests: 4 No. of Tests: 6 or multiple No. of Tests: 2 No. of Tests: 1 No. of Tests: 2 No. of Tests: 1 500VDC through a 130 kOhm limiting resistor. 1, 2 or multiple 1 - 50 mOhm 1 mOhm ±2% of full scale ±1 LSD 51 - 200 mOhm 1 mOhm ±2% of full scale ±1 LSD 0 - 2000 mOhm 1 mOhm ±2% of full scale ±1 LSD 1, 2, or multiple 0 - 1000 mA RMS (@ <250VAC) 1 mA ±2% of full scale ±1 LSD 1 or multiple 1 - 16A RMS (@ <250VAC) 1 mA ±1% of full scale ±1 LSD 1 or multiple

1-2

The following leakage currents are measured for each module: Patient leakage current Mains on applied part leakage current Patient Auxiliary current Substitute patient leakage current Floating dual lead measurement of leakage currents No. of Tests: 6 No. of Tests: 2 No. of Tests: 6 No. of Tests: 1 No. of Tests: Multiple

In one test run a maximum of 11 modules with different protection classes may be tested. 6. Accuracy Range 1 Resolution Accuracy Range 2 Resolution Accuracy Range 3 Resolution Accuracy 7. Frequency Response DC - 1 MHz (-3dB) with a crest factor of >2 The applied test voltage for patient leakage current is 110% of the line voltage, delivered through a limiting resistor of 47 kOhm. 0- 100 µA 1 µA ±2% of full scale ±1 LSD 100- 1000 µA 1 µA ±2% of full scale ±1 LSD 1,0 - 10,0 mA 1 µA ±1% of full scale ±1 LSD

1.3 General Information
Temperature Requirements +15°C to +35°C when operating 0°C to +50°C in storage Display Type Alphanumeric format Data Input/ Output (2)

LCD graphic display 4 lines, 40 characters Parallel printer port (1); Bi-directional RS -232C (1) for computer control From 110 VAC to 240 VAC, 47/ 63 Hz 3900 VA Do not exceed 16 Amps of current at any voltage within the operating range. Installation Category II

Power

Fuses

Two 16 Amp, 250V slow blow fuses

1-3

Mechanical Specifications Housing Height Width Depth Weight

Metal case 13.2 cm 34.2 cm 30.5 cm 5.8 kg 3.9 in. 9.8 in. 11.0 in. 4.1 lbs.

Standard Accessories User and Service Manual QA-90 Additional Accessories Carrying Case Carrying case, ext. printer Bar Code Reader Isolating transformer 400VA Isolating transformer 800VA Test unit (ESA) E input measuring cable (2m) E input measuring cable (5m) Clamp - crocodile type PRO-Soft QA-90 software PRO-Soft QA-90 DEMO User/Service Manual PRO-Soft QA-90

(P/N 11025)

(P/N 11100) (P/N 10500) (P/N 11400) (P/N 11401) (P/N 11410) (P/N 11402) (P/N 11411) (P/N 11415) (P/N 11412) (P/N 11200) (P/N 11201) (P/N 11225)

Storage Store in the carrying case in dry surroundings within the temperature range specified. There are no other storage requirements. Periodic Inspection The unit should be calibrated every 12 months.

1-4

2. Installation
This chapter explains unpacking, receipt inspection and claims, and the general procedures for initial QA-90 setup. Example test setup procedures are contained in Chapter 4, Example Test Measurements. 2.1 Receipt, Inspection and Return 1. 2. Inspect the outer box for damage. Carefully unpack all items from the box and check to see that you have the following items: · · 3. QA-90 Electrical Safety Tester (P.N. 11200) QA-90 User and Service Manual (P.N. 11025)

If you note physical damage, or if the unit fails to function according to specification, inform the supplier immediately. When METRON AS or the company's representative, is informed, measures will be taken to either repair the unit or dispatch a replacement. The customer will not have to wait for a claim to be investigated by the supplier. The customer should place a new purchase order to ensure delivery. When returning an instrument to METRON AS, or the company representative, fill out the address label, describe what is wrong with the instrument, and provide the model and serial numbers. If possible, use the original packaging material for return shipping. Otherwise, repack the unit using: · · · a reinforced cardboard box, strong enough to carry the weight of the unit. at least 5 cm of shock-absorbing material around the unit. nonabrasive dust-free material for the other parts.

4.

Repack the unit in a manner to ensure that it cannot shift in the box during shipment. METRON's product warranty is on page ii of this manual. The warranty does not cover freight charges. C.O.D. will not be accepted without authorization from METRON A.S or its representative.

2-1

2.2 Setup 1. Equipment connection is as shown in the typical setup below. Attach the printer cable to the 25-pin outlet port.

2.

If ansur or PRO-Soft is being used, attach an RS-232C (null modem/data transfer configured) cable to the 9-pin D-sub outlet port located at the rear of the QA-90. Do not attach the printer cable to the QA-90. See below.

NOTE Some RS-232C cables are missing the connection between the seventh and the eighth wires in the cable. The cable may still be called NULL-modem, but it will not work with the QA-90. Refer to the ansur or PRO-Soft QA-90 Users Manual for more information.

2.3 Power Main On/Off Switch. QA-90 should remain off for at least 5 seconds before switching on again, in order to allow the test circuits to discharge fully. 2.4 Ansur & PRO-Soft Ansur and PRO-Soft are front-end test automation and presentation tools for METRON's QA-90 Electrical Safety Analyzer. It allows you to conduct the same tests, but by remote control via an computer running Windows 98, 2000 or XP. Additionally, the program has ad-

2-2

ditional features to automate and enhance your electrical safety testing. Each of the QA-90 tests can be run independently from ansur or PRO-Soft in the "Manual" test mode. Results are shown on the PC screen during testing, and the user is prompted to set the tested equipment accordingly. At the conclusion of tests, the user may print a report, store the test and results on disk, or both. Combinations of tests can be created and stored as "Test Template." The program maintains a library of these templates. In this way you can store and retrieve templates that are appropriate for each kind of equipment being tested at your facility. Templates can include checklists, written procedure, and equipment data in the form of a test "Protocol." The equipment data can be entered manually into the protocol, or it may be retrieved by ansur or PRO-Soft from a database program or other equipment files. Protocols can be created easily for each item of equipment in your inventory, and stored for use. Test protocols with results can be printed, or stored on disk, and the results of testing can be sent back to the equipment database to close a work order and update the service history.

NOTE Ansur and PRO-Soft has its own user manual, which contains all the information concerning the program. If you order a demonstration version of the program you also receive the manual.

2-3

This page intentionally left blank.

2-4

3. Operating QA-90
This chapter explains the operating controls, switches and menus of the QA-90, details how to use them in testing, and provides general information on printouts and operator maintenance. 3.1 Control Switches and Connections Front Panel

Max 265 VAC Max 500 VDC

1.

Key Pad

11 alphanumeric keys, used to enter information. Patient Leads: Clear: Return: Enter: New window for recording patient inputs. Clears the whole display Deletes the last character Records entered data

2.

Function Keys

F1-F4 are used to select the functions shown in the menu bar at the bottom of the display, i.e., for selecting the function that is directly above the key. F5-F7 are used to select the function, or enter information in the message field in the same line. Shows messages, test results and function menus.

3

LCD Display

3-1

4

Patient Leads

For connecting patient inputs. Caution: Max 256 VAC Max 500 VDC can be present during specified tests

5. 6. 7. 8.

Dual Encl. Earth Contact

E+ and E-, floating inputs/outputs. These function in the manner of standard multimeter leads. Enclosure (for connecting to the casing of the instrument under test). Extra earth connection for calibrating measuring leads. For connecting the power plug of the instrument to be tested.

Rear Panel

9. 10. 11. 12. 13. 14. 15. 16.

Power Switch RS-232 Serial Port Bar Code Port Printer Outlet Port Mains QA-90 Auxiliary Power Fuse Earthing Contact

Turns power ON and OFF. 9-pin D-sub 9-pin D-sub. HP-Smartwand Interface (TTL). 25 pin D-sub. Centronic output. Mains connection for test instrument. Auxiliary power connection for instrument under test. Mains fuses 2 x 16 Amps @ 220V Extra earthing point.

3-2

3.2 Key Pad Functions The alphanumeric keys comprise both numbers and letters. Hold the key in and it moves automatically from character to character.

3.3 Function of The Bar Code Reader The bar code reader may be used in the main menu and in the patient lead to record respectively the instrument code and class, module code, number of leads and type. The program will select the correct screen display, depending on the bar code format. The program will give a beep if a wrong format is read.
INSTRUMENT CODE FORMAT
FIELD DESC. Instrument code Separator Instrument class VALUE ACCEPTED alphanumeric + CLI, CL2, I.P MAX. CHARACTERS 20 1 (must be included) 3 (must be included)

Example: abcdefg + CL2

Instrument code = abcdefg Instrument class = CL2
MAX. CHARACTERS 20 1 (must be included) 2 1 (must be included) 2 (must be included)

PATIENT LEAD FORMAT
FIELD DESC. Module code Separator No. of leads Separator Type VALUE ACCEPTED alphanumeric + numeric (0-99) + BF, CF, B

Example: mnopqrst + 2 + B Module code = mnopqrst Number of leads = 2 Module type = B

3-3

3.4 Menu and Function Keys The QA-90 uses displays, function keys and a keypad to provide flexibility and control over operations. The top three lines in the display are used for messages, status and results. The menu bar is shown at the bottom of the display. Function keys are numbered from F1 to F7. A function/menu is selected by pressing that key which is directly below/to the right of the menu unit shown in the display. 3.5 Display Menus and Messages 1. Startup Screen. The following screens will be displayed in sequence for the first 10 seconds after the QA-90 has been switched on.

METRON QA-90 Boot software version: X.XX

F7 F6 F5

F1

F2

F3

F4

METRON QA-90 Version: X.XX Date: DD/MM/YYYY

F7 F6 F5

F1

F2

F3

F4

METRON QA-90 Electrical Safety Analyzer TOTAL MEMORY : 114688 TOTAL MEMORY USED : 6884 TOTAL FREE SPACE : 107804 F1 F2 F3 F4

F7 F6 F5

3-4

2.

Main Menu
Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4 F7 F6 F5

3. MORE (F1). When MORE (F1) is pressed, the following display will appear:
Test according to Test Type Test Mode GO BACK F1 F2 : IEC60601.1 ......> : Rapid ...........> : Automatic .......> START F3 F4 F7 F6 F5

This window offers the following functions: · Press Test according to (F7) to select either IEC 60601.1, IEC 60601.2.4, IEC 61010-1, UL 2601.1, IEC61010, AS 3200.1, HEI 95, DB 9801, ANSI/AAMI, VDE 0751:101990 or VDE 0751:10-2001 Standards. Press Test Type (F6) to select either Rapid or Normal test type. Press Test Mode (F5) to choose between Automatic and Manual test. Press GO BACK (F2) to return to the previous display. Press START (F4) to start the test.

· · · · 4.

MEMORY (F2) QA-90's memory is divided into two parts: tests and sequences. If you have several equivalent instruments to test, you can define one test sequence for all the instruments and store it as a sequence. You use this sequence to test all the instruments and store each of the test results with the instrument's respective equipment code. The functions in the MEMORY menus enable you to store, retrieve, transfer, print and delete test results and sequences from the memory. The four main memory displays are as follows.

3-5

Equipment Code : ...> Store test in memory ..................> Display test results ..................> MORE SEQ.MEM TEST MAIN MENU F1 F2 F3 F4

F7 F6 F5

Equipment Code : ...> Recall and transfer test from memory ..> Recall and print test from memory .....> MORE GO BACK MAIN MENU F1 F2 F3 F4

F7 F6 F5

Transfer all tests from memory ........> List test in memory to printer ........> Print all tests in memory .............> MORE GO BACK MAIN MENU F1 F2 F3 F4

F7 F6 F5

Equipment Code : ...> Delete test in memory .................> Delete all tests in memory ............> MORE GO BACK MAIN MENU F1 F2 F3 F4

F7 F6 F5

The F5 and F6 function keys generate new displays as confirmation of an executed function or error message. · · Press GO BACK (F2) to return to the previous display. Press MAIN MENU (F4) to start the test.

The three main sequence displays are shown below.
Sequence name : ...> Store seq. in memory ..................> Print sequence ........................> MORE GO BACK SEQ MAIN MENU F1 F2 F3 F4 F7 F6 F5

3-6

List all sequences in memory ..........> Print all sequences in memory .........> MORE F1 GO BACK F2 MAIN MENU F3 F4

F7 F6 F5

Sequence name : ...> Delete sequence in memory .............> Delete all sequences in memory ........> GO BACK MAIN MENU F1 F2 F3 F4

F7 F6 F5

5. SETUP (F3) This function is used for entering general information in connection with the test. Seven main displays are shown below.

SYSTEM SETUP Operator : Establishment : MORE STORE CAL. F1 F2

F7 ...> ...> MAIN MENU F3 F4 F6 F5

QA-90 Serial no IT-Net Power-up delay time MORE GO BACK F1 F2

: XXXXX .> : N ..........> : 2 Seconds.> CAL. MAIN MENU F3 F4

F7 F6 F5

QA-90 Serial no IT-Net Power-up delay time MORE GO BACK F1 F2

: XXXXX .> : N ..........> : 2 Seconds.> CAL. MAIN MENU F3 F4

F7 F6 F5

3-7

Stop at new power config : Stop before new power config : Stop at new module : MORE GO BACK CAL. MAIN F1 F2 F3

N ...> N ...> N ...> MENU F4

F7 F6 F5

Multiple Protective Earth Tests : Multiple Enclosure Tests : PE Test Current : MORE GO BACK CAL. MAIN F1 F2 F3

N ...> N ...> 25 A.> MENU F4

F7 F6 F5

External Isolating Transformer : Acoustic alarm for failed test : Auxiliary Power : MORE GO BACK CAL. MAIN F1 F2 F3

N ...> Y ...> N ...> MENU F4

F7 F6 F5

Language : English ....................> Date : XX/XX/XXXX (DD/MM/YYYY) ...> Time : XX:XX:XX (HH:MM:SS) .....> MORE GO BACK AM/PM MAIN MENU F1 F2 F3 F4

F7 F6 F5

Report page length(no of lines) :58 ...> FIRMWARE Ver. XX.XX GO BACK F1 F2

F7 F6 F5

CAL.

MAIN MENU F3 F4

6. Recording and Storing Patient Modules and Patient Leads. · Press PL on the keypad to obtain the window for recording patient modules and leads.
F7 F6 F5

Module Code : .....> No of leads : .......................> Type : CF.......................> ADD GO BACK PREV. NEXT F1 F2 F3 F4

3-8

· · · · · · · · ·

Go to the next patient module stored Go to the previous patient module stored. Return to the previous menu Press ADD (F1) to save in the memory Press Module Code (F7) to record the code/name of a new patient module. Press ENTER (). Press No of leads (F6) to enter the number of patient leads. Press ENTER (). Press Type (F5) to select protective class (B, BF, CF). Press ADD (F1) to save in the memory. The number of modules stored will be shown in parenthesis in the Module Code line. Repeat the above guidelines to enter the next module. Press NEXT (F4) to go to the next patient module stored. Press PREV (F3) to go to the previous patient module stored. Press GO BACK (F2) to return to the previous menu.

· · · ·

7. START (F4). This function starts the test sequence. Manual or automatic test sequences are selected under MORE (F1) in MAIN MENU. 3.6 Measurements with Several Modules in Manual Mode When you perform measurements on several modules in Manual Mode and want to select which module to measure, use the following procedure: 1. Press MORE (F1) in MAIN MENU.
Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4 F7 F6 F5

3-9

2.

Press Test Mode (F5) and select Manual. Press START (F4).
Test according to Test Type Test Mode GO BACK F1 F2 : IEC60601.1 ......> : Rapid ...........> : Manual ..........> START F3 F4 F7 F6 F5

3.

Press MORE (F1) in repeatedly until you find the desired measurement, e.g., Mains on applied part (F5).
MANUAL TEST SETUP Mains Voltage .........................> Current Consumption ...................> MORE MAIN MENU F1 F2 F3 F4 F7 F6 F5

4.

Press PL on the keypad to obtain the recording window for Module Code and No of leads.
Module Code : .....> No of leads : .......................> Type : CF.......................> ADD GO BACK PREV. NEXT F1 F2 F3 F4 F7 F6 F5

5. 6. 7. 8. 9.

Select which Module Code to be measured with NEXT (F4) or Prev. (F3). Press GO BACK (F2). Press START (F1) to start the test. Press STOP (F1) to stop the test. To select a new module, press PL and repeat the same procedure.

3.7 Storing Setup Parameters in Flash Memory The following parameters can be stored in Flash memory: · · Operator Establishment

3-10

· · ·

Serial Number Language Calibration parameters

To store the setup parameters in Flash memory, go to SETUP (F3) in MAIN MENU.
Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4 F7 F6 F5

Press STORE (F2) in SYSTEM SETUP and the display will show you the setup parameters stored in flash.
SYSTEM SETUP Operator : Establishment : MORE STORE CAL. F1 F2 F7 ...> ...> MAIN MENU F3 F4 F6 F5

3.8 Upgrading the QA-90 Software Program For installing software, version 2.00 or higher, use the following procedure: 1. Preparation
IMPORTANT! When upgrading the software on calibrated units, you should note the calibration constants in the unit. There are seven constants that can be read on the unit display.

· ·

Turn on the unit. Press SETUP (F3) from the Main menu.
F7 F6 F5

Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4

3-11

·

Press CAL (F3) from the System Setup.
F7 ...> ...> MAIN MENU F3 F4 F6 F5

SYSTEM SETUP Operator : Establishment : MORE STORE CAL. F1 F2

·

Then go to the hidden menu under (F1), in the menu for Self Calibration.
F7 F6 F5

SELF CALIBRATION Calibrate test lead, enclosure/earth ..> Calibrate test lead, dual float .......> GO BACK MAIN MENU F1 F2 F3 F4

· · ·

Note the values on the three constants that appear on the display. Press MORE (F1) to get the three next constants. Press MORE (F1) to get the last constant. NB: Automatic MAP calibration: ONLY for production calibration. The QA-90 software upgrade contains two files: comqa90.exe, the communication between computer and the QA-90, and; qa90-xxx.a07 is the program for QA-90. (XXX indicates the version of the firmware.)

2. Prepare the QA-90 for the software upgrade. · · · Reading the program to Flash memory in the QA-90. Press and hold PL and CLR at the same time while turning the QA-90 on. The display will show the following:

The software in the QA-90 is missing. To download new software, run COMMQA90 on the PC and press ENTER on QA-90

3-12

· · · · · · ·

The QA-90 is ready to receive data. Connect an RS-232 cable between PC and QA-90. Insert the disk that contains the QA-90 upgrade program. Write COMMQA90.EXE (DOS) Press ENTER () Choose menu 1 SET PARAMETERS and choose new parameters or default values. Choose menu 2 START COMMUNICATION by pressing 2. Enter filename QA90-XXX.A07 where XXX is the version of the firmware. Check the file. Press ENTER () on the computer. Press ENTER () on the QA-90 and the transmission will start.

3. Prepare the PC for the upgrade.

A menu will appear on the computer:

· ·

If the communication is successful, >>>>>>> will appear continuously at the display on the QA-90. When the program is transmitted, a normal startup menu will appear on the QA-90. The computer will either show a picture for program transmission, or a clear screen. 4. Check the Calibration Constants Check the calibration constants in the QA-90 through the hidden menu. If necessary, press F5, F6 or F7 to type your noted calibration constants, then press ENTER (). To store in Flash memory, go back to SYSTEM SETUP.
SYSTEM SETUP Operator : Establishment : MORE STORE CAL. F1 F2 F7 ...> ...> MAIN MENU F3 F4 F6 F5

Press STORE (F2) and the display will show you setup parameters stored in flash.

3-13

3.9 Cleaning the QA-90 The outside of the instrument may be cleaned using a damp cloth and mild detergent. Please note that solvents like Methanol may dammage the overlay and cabinet.

3-14

This page intentionally left blank.

3-15

4. Example Test Measurements
This chapter contains test examples for the QA-90, illustrating equipment connections for the tests, as well as step-by-step procedures for obtaining desired test measurements. For more information on safety testing, and an explanation of protective classes, refer to Appendix A, IEC 60601.1, UL 2601.1 and VDE 0751 Testing. 4.1 Test Lead Calibration The self-calibration function of the QA-90 is used to determine test lead resistance, and for the values to be taken into account during subsequent testing. 1. Prior to performing self-calibration connect a test lead between the enclosure and earth connectors, or the dual lead inputs on the front face of the QA-90 (see below). Disconnect all other leads.

2. 3. 4.

Press SETUP (F3) in MAIN MENU. Press CAL (F3) in SYSTEM SETUP. In the SELF CALIBRATION window select an option by pressing either Calibrate test lead, enclosure/ground (F6) or Calibrate test lead, dual float (F5).
Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4 F7 F6 F5

4-1

SYSTEM SETUP Operator : Establishment : MORE STORE CAL. F1 F2

F7 ...> ...> MAIN MENU F3 F4 F6 F5

SELF CALIBRATION Calibrate test lead, enclosure/earth ..> Calibrate test lead, dual float .......> GO BACK MAIN MENU F1 F2 F3 F4

F7 F6 F5

5.

The test result for the calibration appears in the display when the test is complete.
Test lead enclosure/grnd Test lead, dual float GO BACK F1 F2 PRINT : : XXX mOhm XXX mOhm MAIN MENU F3 F4 F7 F6 F5

4.2 Connecting an Instrument without Patient Inputs Connect the mains plug of the instrument to be tested to the QA-90's terminal on the front panel. Then, connect a calibrated test lead between the ENCL connector on the front panel of the QA-90 and the casing of the instrument to be tested. (See below)

NOTES Ensure that the mains switch on the instrument to be tested is switched ON. If the message REVERSED POLARITY is shown in the MAIN MENU, reverse the mains plug.

1.

Press Equipment Classification (F6) and select classification.

4-2

2. 3. 4. 5. 6.

Press MORE (F1). Press Test according to (F7) and select test standard. Press Test Type (F6) and select either Rapid or Normal. Press Test Mode (F5) and select either Automatic or manual. Press START (F4) to start the test.
F7 F6 F5

Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4

Test according to Test Type Test Mode GO BACK F1 F2

: IEC60601.1 ......> : Rapid ...........> : Automatic .......> START F3 F4

F7 F6 F5

4.3 Connecting an Instrument with Patient Inputs This includes IEC Classifications Body (B), Body Float (BF), and Cardiac Float (CF). Connect the mains plug of the instrument to be tested to the QA-90's terminal on the front panel. Then, connect a calibrated test lead between the ENCL connector on the front panel of the QA-90 and the casing of the instrument to be tested. Following that, connect patient lead between the instrument to be tested and Patient leads on the QA90. (See below)

NOTE Ensure that the mains switch on the instrument to be tested is switched ON.

4-3

1. 2. 3. 4. 5. 6. 7. 8. 9.

Press Equipment Classification (F6) and select classification. Press MORE (F1). Press Test according to (F7) and select test standard. Press Test Type (F6) and select either Rapid or Normal. Press Test Mode (F5) and select either Automatic or Manual. Press PL on the keypad to obtain the recording window for patient lead. Press Module Code (F7) if you wish to record the name of a new module. Press ENTER (.). Press No of leads (F6) to enter the number of leads.

10. Press ENTER (.). 11. Press Type (F5) to select protective class. 12. Press ADD (F1) to accept. The number of modules stored will be shown in parenthesis in the Module Code line. 13. Repeat, as required, to enter a new patient module. 14. Press GO BACK (F2) to return to the menu. 15. Press START (F4) in the MAIN MENU to start the test.
Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4 F7 F6 F5

Test according to Test Type Test Mode GO BACK F1 F2

: IEC60601.1 ......> : Rapid ...........> : Manual ..........> START F3 F4

F7 F6 F5

4-4

Module Code : .....> No of leads : .......................> Type : CF.......................> ADD GO BACK PREV. NEXT F1 F2 F3 F4

F7 F6 F5

4.4 Power Cable Test The earth lead in the power cable is tested as follows. Plug the power cable into the front panel of the QA-90, then connect a calibrated test lead between ENCL. and the each pin on the power cable.

NOTE Ensure that the test leads are calibrated BEFORE the test.

1. 2. 3. 4. 5. 6.

Press MORE (F1) in MAIN MENU. Press Test Mode (F5) and select Manual. Press START (F4). Press MORE (F1) in MANUAL TEST SETUP. Press Protective Earth (F7). Press START (F1).
F7 F6 F5

Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4

Test according to Test Type Test Mode GO BACK F1 F2

: IEC60601.1 ......> : Rapid ...........> : Automatic .......> START F3 F4

F7 F6 F5

4-5

MANUAL TEST SETUP Mains Voltage .........................> Current Consumption ...................> MORE MAIN MENU F1 F2 F3 F4

F7 F6 F5

Protective Earth ......................> Insulation Resistance .................> Earth Leakage Current .................> MORE GO BACK MAIN MENU F1 F2 F3 F4

F7 F6 F5

Time : Result : Limit : START F1 mOhm mOhm GO BACK F2

Seconds

F7 F6 F5

PRINT

MAIN MENU F3 F4

4.5 Current Measurement Test (Dual Lead) This test measures leakage current from one instrument to another instrument.

1. 2. 3. 4.

Press MORE (F1) in MAIN MENU. Press Test Mode (F5) and select Manual. Press START (F4). Press MORE (F1) in MANUAL TEST SETUP.

4-6

5. 6. 7.

Press MORE (F1) three more times. Press Current Measurement Dual Lead (F7). Press START (F1).
F7 F6 F5

Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4

Test according to Test Type Test Mode GO BACK F1 F2

: IEC60601.1 ......> : Rapid ...........> : Automatic .......> START F3 F4

F7 F6 F5

MANUAL TEST SETUP Mains Voltage .........................> Current Consumption ...................> MORE MAIN MENU F1 F2 F3 F4

F7 F6 F5

Current Measurement Dual Lead .........> Voltage Measurement Dual Lead .........> Resistance Measurement Dual Lead ......> GO BACK START F1 F2 F3 F4

F7 F6 F5

Open Supply ...........................> Result : µA Measure between E+ and E- inputs START GO BACK PRINT MAIN MENU F1 F2 F3 F4

F7 F6 F5

4-7

4.6 Voltage Measurement Test (Dual Lead) Measuring voltage potentials to a specified reference.

1. 2. 3. 4. 5. 6. 7.

Press MORE (F1) in MAIN MENU. Press Test Mode (F5) and select Manual. Press START (F4). Press MORE (F1) in MANUAL TEST SETUP. Press MORE (F1) three more times. Press Voltage Measurement Dual Lead (F6). Press START (F1).
F7 F6 F5

Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4

Test according to Test Type Test Mode GO BACK F1 F2

: IEC60601.1 ......> : Rapid ...........> : Automatic .......> START F3 F4

F7 F6 F5

MANUAL TEST SETUP Mains Voltage .........................> Current Consumption ...................> MORE MAIN MENU F1 F2 F3 F4

F7 F6 F5

4-8

4-9

Current Measurement Dual Lead .........> Voltage Measurement Dual Lead .........> Resistance Measurement Dual Lead ......> GO BACK START F1 F2 F3 F4

F7 F6 F5

Voltage :

V

F7 F6 F5

Measure between E+ and E- inputs START GO BACK PRINT MAIN MENU F1 F2 F3 F4

4.7 Resistance Measurement (Dual Lead) Measuring protective earth on fixed installations.

1. 2. 3. 4. 5. 6. 7.

Press MORE (F1) in MAIN MENU. Press Test Mode (F5) and select Manual. Press START (F4). Press MORE (F1) in MANUAL TEST SETUP. Press MORE (F1) three more times. Press Resistance Measurement Dual Lead (F5). Press START (F1).
F7 F6 F5

Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4

4-10

Test according to Test Type Test Mode GO BACK F1 F2

: IEC60601.1 ......> : Rapid ...........> : Automatic .......> START F3 F4

F7 F6 F5

MANUAL TEST SETUP Mains Voltage .........................> Current Consumption ...................> MORE MAIN MENU F1 F2 F3 F4

F7 F6 F5

Current Measurement Dual Lead .........> Voltage Measurement Dual Lead .........> Resistance Measurement Dual Lead ......> GO BACK START F1 F2 F3 F4

F7 F6 F5

Open Supply ...........................> Result : µA Measure between E+ and E- inputs START GO BACK PRINT MAIN MENU F1 F2 F3 F4

F7 F6 F5

4.8 Connecting Auxiliary Power Source / Isolating Transformer External power cable gives power out on the contact on the front panel to the instrument under test.

NOTE The QA-90 power cable must also be connected.

4-11

If the equipment under test is to be tested for voltages and/or frequencies that differs from the nominal mains supply, the test voltage must be connected to the auxiliary inlet. To route the auxiliary power to the contact on the front panel, the procedure below must be executed. 1. Press SETUP (F3) in MAIN MENU. 2. 3. 4. Press MORE (F1) in SYSTEM SETUP. Press MORE (F1) three more times. Press External Isolating Transformer (F7) and select Y (Yes).
F7 F6 F5

Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4

SYSTEM SETUP Operator : Establishment : MORE STORE CAL. F1 F2

F7 ...> ...> MAIN MENU F3 F4 F6 F5

External Isolating Transformer : Acoustic alarm for failed test : Auxiliary Power : MORE GO BACK CAL. MAIN F1 F2 F3

Y ...> Y ...> Y ...> MENU F4

F7 F6 F5

4-12

This page intentionally left blank.

4-13

5. Control and Calibration
This chapter explains the QA-90 maintenance procedures, including testing and calibration. 5.1 Required Test Equipment · · · · · · · · 5.2 Preparation and Adjustments
IMPORTANT! BEFORE TESTING THE QA-90 MUST BE POWERED ON FOR AT LEAST 30 MINUTES PRIOR TO CONDUCTING ANY TESTING.

Power source with a range of 0 - 10 mA (Variref VF12). Digital multimeter: Range 0 - 16 A. Digital multimeter: Range 0 V and 500 V (HP34401 or similar). Metron calibrating unit for earth protection, M. Metron calibrating unit for insulation resistance, M. Adjustable equipment with an effective output ranging from 0 to 16A for calibrating current consumption. Test leads. Insulated transformer.

1.

Startup Preparation a. b. c. d. e. Press MORE (F1) in MAIN MENU. Press Test according to (F7), and select UL 2601.1. Press Test Type (F6), and select Rapid. Press Test Mode (F5), and select Manual. Press START (F4) then press MORE (F1) twice to advance to Manual Test Setup Window 3.
F7 F6 F5

Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4

5-1

Test according to Test Type Test Mode GO BACK F1 F2

: IEC60601.1 ......> : Rapid ...........> : Automatic .......> START F3 F4

F7 F6 F5

MANUAL TEST SETUP Mains Voltage .........................> Current Consumption ...................> MORE MAIN MENU F1 F2 F3 F4

F7 F6 F5

Protective Earth ......................> Insulation Resistance .................> Earth Leakage Current .................> MORE GO BACK MAIN MENU F1 F2 F3 F4

F7 F6 F5

Enclosure Leakage Current .............> Patient Leakage Current AC ............> Patient Leakage Current DC ............> MORE GO BACK MAIN MENU F1 F2 F3 F4

F7 F6 F5

2.

Patient Lead Definition. To facilitate some calibration measurements patient leads need to be defined. To do so: a. Press the PL key on the keypad, and the following window appears:
F7 F6 F5

Module Code : .....> No of leads : .......................> Type : CF.......................> ADD GO BACK PREV. NEXT F1 F2 F3 F4

NOTE The module numbers do not need to be entered, as they are defaulted as (1, 2, etc.)

b. c.

For Patient Leads 1-10 (white leads) press No. of leads (F6) and key in "10". Press the Enter key. Press ADD (F1), then NEXT (F4).

5-2

d. e.

For Patient Lead 11 (blue lead) press No. of leads (F6) and key in "1". Press the Enter key. Press ADD (F1), then GO BACK (F2) to return to Manual Test Setup Window 3.

3. Display Contrast Adjustment a. b. Remove the case housing. Set the contrast in the display by adjusting the potentiometer located to the left of the CPU.

4. Offset Adjustments.1 Before conducting measurements offset adjustments must be made to the measuring devices. (Refer to schematic diagram E160.20.2600.U3, Prog. Amplifier/Lowpass QA90) a. RMS-DC Converter. The RMS-DC converter output offset may be adjusted if S2630 pins 1 and 2 are shunted. If S2630 pins 2 and 3 are shunted, the offset of U2630 is disabled. Never move this jumper on a calibrated QA-90, because it will cause a fault of about 1.5% on the measurements. However, if the QA-90 is not yet calibrated, the jumper may be moved. The offset adjustment of U2630 is not mandatory, but it results in small measuring improvements. Before offset adjustment we recommend you remove measuring amplifier U2600 from its socket. Make a short between SGND and pin 1 of U2630. Measure the voltage of output pin 6 and SGND. Adjust R2630 to an offset value as close to zero as possible. b. Measuring Amplifier. Adjustment is possible if S2600 pins 2 and 3 are shunted. The adjustment is disabled if the jumper is between pins 1 and 2. 1) Select Enclosure Leakage Current (F7). 2) Select Normal Condition (F7). 3) Press START (F1) and generate a current of 1 mA between ENCL. and Chassis. Note the result and limits readout on the QA-90's display. 4) Change polarity and take a new measurement. Adjust values for both polarities to be equal. Several adjustments in both directions may have to be made.
1

The offset adjustment capability is unavailable in QA-90 firmware versions before 2.xx.

5-3

5) Select GO BACK (F2) to return to Manual Test Setup Window 3.

WARNING! HIGH VOLTAGES ARE CAPABLE OF CAUSING DEATH! USE EXTREME CAUTION WHEN PERFORMING TESTS AND CALIBRATION. USE ONLY INSULATED TOOLS WHEN THE UNIT IS PLUGGED IN, AND THE CASE HOUSING IS OFF.

5.3 Calibration 1. Enclosure Leakage Current a. b. c. d. e. Select Enclosure Leakage Current (F7). Select Normal Condition (F7). Press START (F1) and generate a current (10 µA - 10 mA) between ENCL. and Chassis. Then, press STOP (F1). Check that both the level of this current, and the readout on the QA-90 display, are equal. Press GO BACK (F2) twice to go to Manual Test Setup Window 2. Press Earth Leakage Current (F5). Select Normal Condition (F7). Press START (F1) and generate a current (10 µA - 10 mA) between EARTH and Chassis. Then, press STOP (F1). Check that both the level of this current, and the readout on the QA-90 display, are equal. Press GO BACK (F2), then MORE (F1) to advance to Manual Test Setup Window 3.
F7 F6 F5

2. Earth Leakage Current a. b. c. d. e.

Enclosure Leakage Current .............> Patient Leakage Current AC ............> Patient Leakage Current DC ............> MORE GO BACK MAIN MENU F1 F2 F3 F4

5-4

3. Patient Leakage Current AC a. b. c. Press Patient Leakage Current AC (F6). Select Normal Condition (F7). Press the PL key on the keypad to ensure that the module with Patient Leads 1-10 (white leads) is active. If not, press PREV (F3) or NEXT (F4) so that it is active. Press GO BACK (F2) to return to the test window, then press START (F1). Use a current source to generate an AC current between Patient Leads 1-10 and EARTH. Check that both the level of this current, and the current readout on the QA-90 display, are equal. Press STOP (F1). Press the PL key on the keypad, then , press PREV (F3) or NEXT (F4) to ensure that the module with Patient Lead 11 (blue lead) is active. Press GO BACK (F2) to return to the test window, then press START (F1) and run the same test between Patient Lead 11 and EARTH. Repeat the test with different current values. Press STOP (F1). Press GO BACK (F2) to return to Manual Test Setup Window 3. Press Patient Leakage Current DC (F5). Select Normal Condition (F7). Press the PL key on the keypad to ensure that the module with Patient Leads 1-10 (white leads) is active. If not, press PREV (F3) or NEXT (F4) so that it is active. Press GO BACK (F2) to return to the test window, then press START (F1). Use a current source to generate an DC current between Patient Leads 1-10 and EARTH. Check that both the level of this current, and the current readout on the QA-90 display, are equal. Press STOP (F1). Press the PL key on the keypad, then , press PREV (F3) or NEXT (F4) to ensure that the module with Patient Lead 11 (blue lead) is active. Press GO BACK (F2) to return to the test window, then press START (F1) and run the same test between Patient

d. e.

f.

g.

h.

4. Patient Leakage Current DC a. b. c.

d. e.

f.

g.

5-5

Lead 11 and EARTH Repeat the test with different current values. Adjust the General I-constant if the values between UUT and reference are not equal. Press STOP (F1). h. Press GO BACK (F2) until you return to Manual Test Setup Window 3. Then, press MORE (F1) to advance to Manual Test Setup Window 4.
F7 F6 F5

Patient Auxiliary Current AC ..........> Patient Auxiliary Current DC ..........> Mains on Applied Part .................> MORE GO BACK MAIN MENU F1 F2 F3 F4

5. Patient Auxiliary Current DC a. b. c. Press Patient Auxiliary Current DC (F6). Select Normal Condition (F7). Press the PL key on the keypad to ensure that the module with Patient Leads 1-10 (white leads) is active. If not, press PREV (F3) or NEXT (F4) so that it is active. Press GO BACK (F2) to return to the test window, then press START (F1). Use a current source to generate an DC current (20 µA) between Patient Leads 1-9 and Lead 10, and between Patient Leads 1-8 and Lead 9. Check that both the level of this current, and the current readout on the QA-90 display, are equal. Press STOP (F1). Press GO BACK (F2), then press MORE (F1) to advance to Manual Test Setup Window 5.
F7 F6 F5

d. e.

f.

Current Measurement Dual Lead .........> Voltage Measurement Dual Lead .........> Resistance Measurement Dual Lead ......> GO BACK START F1 F2 F3 F4

6. Current Measurement Dual Lead a. b. c. Press Current Measurement Dual Lead (F5). Select Normal Condition (F7). Press START (F1).

5-6

d. e. f. g. a. b. c. d. e. f.

Use a current source to generate a current between the DUAL (Red and Black) inputs. Press STOP (F1). Check that both the level of this current, and the current readout on the QA-90 display, are equal. Press GO BACK (F2) to Manual Test Setup Window 5. Press Voltage Measurement Dual Lead (F7). Select Normal Condition (F7). Press START (F1). Use a power/voltage source to generate an AC voltage. Press STOP (F1). Check that the level of this current, and the current readout on the QA-90's display are equal. If necessary, adjust the voltage to be equal on UUT and reference. The constant used is the SP/VDML U-constant. Repeat the measurement, but with DC voltage. Press GO BACK (F2) four times to go to Manual Test Setup Window 1.

7. Voltage Measurement Dual Lead

g. h. 8.

Supply Voltage. An insulated transformer is used for this measurement, except for step 7.f. below. a. b. c. d. e. f. g. Press Supply Voltage (F6). Check the voltage between LIVE and NEUTRAL on the front panel terminals with a calibrated instrument. Select Live - Neutral (F7) and press START (F1). Press STOP (F1). Check the voltage. Select Neutral - Ground (F7) and press START (F1). Press STOP (F1). Check the voltage. Connect a variac to the terminals on the front panel and check the voltage. The Live - Neutral setting is used to measure, for example, 70 and 250 VAC. It is not necessary to use an insulated transformer. Press GO BACK (F2) to return to Manual Test Setup Window 1.

h.

5-7

9.

Current Consumption. This measurement is done with a resistive load connected to the mains outlet on the front panel of the QA-90. The reference digital multimeter (HP34401 or similar), in current (AC) mode, is in series with the load. Use the digital multimeter in the 0 - 3 A range, and the same instrument with a current probe (HP34330A or similar) in the range > 3 A. The probe and digital multimeter are calibrated together. a. b. c. Press Current Consumption (F5). Press START (F1). Press STOP (F1). Check that the UUT and the reference digital multimeter values are equal. If they are not equal, adjust the CC I-constant. Press GO BACK (F2) , then press MORE (F1) to advance to Manual Test Setup Window 2.

d.

10. Protective Earth
IMPORTANT ENSURE THAT THE TEST LEADS ARE CALIBRATED BEFORE CONDUCTING THIS MEASUREMENT. METRON RECOMMENDS THAT, WHEN CALIBRATING THE QA-90, TEST LEAD CALIBRATION BE DONE AT LEAST TWICE.

a. b. c.

Press Protective Earth (F7). Connect the test leads between EARTH and ENCL. Press START (F1). QA-90 will time the measurement and stop automatically. Take measurements at 0, 100, 500, 1000 and 2000 m with a Metron calibrating unit. Note that the test current is delivered by a transformer located on the right side of the cabinet. In low mains voltage, this transformer will not deliver the IEC/UL-recommended 10 A. If you are doing repetitive testing and the transformer gets hot, it may turn off for a while to prevent overheating. It is not rated to deliver in excess of 25 A in short circuit condition. A transformer for 100-120 V mains is available from METRON. Press GO BACK (F2) to return to the previous display.

d.

11. Insulating Resistance. Insulation resistance measurements are done in both the Mains to Case and Applied Part to Case modes. a. b. Ensure the QA-90 is ON for at least 30 minutes. Press Insulating Resistance (F6).

5-8

c. d.

Select Mains To Case (F7). Connect a voltmeter or a digital multimeter, with an input impedance of 10 M, between ENCL. and the mains outlet on the front panel. Press START (F1). Measure the test voltage. Trim the test voltage to within 500 ± 1 V. Then, use a calibrated resistor to measure values in the range 1-200 M. Using a reference resistor of 90-110 M, calculate the IR Uconstant2 such that the QA-90 measures as closely as possible to that reference. Using a reference resistor of 190-210 M, calculate the IR R-constant such that the QA-90 measures as closely as possible to that reference.
IMPORTANT YOU HAVE TO PERFORM SEVERAL CALCULATIONS OF THESE CONSTANTS BECAUSE OF THEIR INFLUENCE ON EACH OTHER. IF THEY ARE CLOSE TO DEFAULT, FEW CALCULATIONS ARE NECESSARY. HOWEVER, IF THE IR-R CONSTANT HAS TO BE HIGH, MEASUREMENTS MUST BE REPEATED TO OBTAIN THE BEST POSSIBLE LINEARITY.

e. f.
NOTE Never set any of the constants higher than 99.999. g.

h.

i. j. k. l.
NOTE Remember that the patient leads must be defined (1-10).

If you are unable to justify levels above 210 M, it may be necessary to change the measurement amplifier. Check the range of 1-200 M (e.g. 1, 10, 100 and 200 M) with Metron's test instrument. Press STOP (F1). Press GO BACK (F2) to return to Main Setup Window 2 and press Mains on Applied Parts (F6). Select (F7) to display Applied Part to Case. Measure the voltage between EARTH and applied part. Adjust the voltage to 253 VAC for European Models or 132 VAC for US Models (Patient Leads 1-10). Check that the current displayed on the UUT is in accordance with Ohm's Law (253 or 132 VAC /test-box imped-

m. Define the Patient Leads 1-10. n.

o.

2

On older QA-90 firmware versions there is only one constant IR U-constant for insulation resistance measurements. On version series 0.xx there are six constants, while on version series 1.xx there are seven. Also with the 1.xx series, the IR U- and IR R-constants are used for insulation resistance measurements.

5-9

ance). The measured alternating voltage, divided by the voltmeter's internal resistance. p. 5.4 Other Calibration Procedures 1. Bar Code Reader (Optional). This checks the QA-90 Bar Code Reader function and its ability to properly accept scan and convert the bar codes into machine-readable format. (See paragraph 3.3 for more information on this feature.) a. b. Attach the scanning wand's TTL interface to the 9-pin Dsub port on the QA-90's rear panel. Return to the Main Menu.
F7 F6 F5

Press GO BACK (F2) to return to the previous display.

Equipment Code : ...> Sequence Name : ...> Class: CL1 ,0 leads in 0 modules.......> MORE MEMORY SETUP START F1 F2 F3 F4

c. d.

Scan an improperly formatted instrument bar code. The QA90 should beep to indicate this error. Scan a properly formatted instrument bar code. Check to see that the Equipment Code and Equipment Class are entered in the display. Follow the procedure in paragraph 3.6 to advance to the display window containing the Module Code, No. of Leads and Type.
F7 F6 F5

e.

Module Code : .....> No of leads : .......................> Type : CF.......................> ADD GO BACK PREV. NEXT F1 F2 F3 F4

f. g.

Scan an improperly formatted patient lead bar code. The QA-90 should beep to indicate this error. Scan a properly formatted patient lead bar code. Check to see that the Module Code, No. of Leads and Type are entered in the display.

5-10

2.

Automatic Test Sequence with Printout. This checks the QA90's sequence testing function. For the printout function a printer must be connected to the QA-90, and be ON. See chapter 2 for additional information. a. Return to the QA-90's main menu, then press MORE (F1) twice to arrive at page 3.
: IEC60601.1 ......> : Rapid ...........> : Manual ..........> START F3 F4 F7 F6 F5

Test according to Test Type Test Mode GO BACK F1 F2

b. c.

Select Automatic in Test Mode (F5). Press START (F4). QA-90 will then commence an automated sequence test. When the test is completed the following Test Results Window appears.
F7 F6 F5

---TEST FAILED--Print Test Results ....................> Print Failed Results ..................> SHOW TEST SHOW FAILED STORE MAIN MENU F1 F2 F3 F4

d.

Press Print test results (F6). Check the printout for legibility. Then, press Print failed results (F5). Check the printout for legibility.

5-11

IMPORTANT ENSURE THAT YOU SAVE THE CALIBRATION CONSTANTS TO FLASH.

5.5 Calibration Constants 1.

SP/VMDL U-constant. Adjusts the voltage measurement for Supply Voltage and Voltage Measurement Dual Lead. When the parameter value increases, the measurement value decreases. Calibrates at 200 - 230 V alternating voltage. This is best done at Dual Lead. Measure with AC voltage with polarity changed. Add the measurements to get the average and calculate the correct constant. PEIRMDL U-constant. Adjust the voltage measurement for Protective Earth and Resistance Measurement Dual Lead. When the parameter value increases, the measurement value decreases. This value does not usually change. The default value is 84.30. IR U-constant. Adjust the resistance measurement for insulating resistance. When the parameter value increases, the measurement value increases. The calibration occurs with a calibrating resistance of 90 - 110 m. Ensure the power of 500 V is set correctly, and that the QA-90 has reached its operating temperature. General I-constant. Adjust the current measurement for PLC, ALC, MAP, EnLC, ELC, CMDL and the IR tests. When the parameter value increases, the measurement value decreases. The calibration and calculation of the parameter occurs with a current of 1000 10000 µA from the calibration instrument. PE/RMDL I-constant. Adjust the current measurement for the PE and RMDL tests. When the parameter value increases, the measurement value increases. CC I-constant. Adjust the current measurement for Current Consumption. When the parameter value increases, the measurement value decreases. IR R-constant. Adjust the resistance measurement for Insulating resistance. When the parameter value increases, the measurement value decreases. The calibration occurs with a calibrating resistance of 190 - 210 m. Make sure that the power of 500 V is set correctly and that the apparatus has reached its operating temperature.

2.

3.

4.

5.

6.

7.

5-12

6. Theory of operation
This chapter provides a detailed description of the functions of the main components of the QA-90, as well as a parts list for crossreference. The QA-90 comprises three printed circuit boards: a metering board, processor and keypad board. The circuit boards are supplied from a 40 W "medical grade" power source, which delivers +12 Volt and +5 Volt. In addition, a temperature-protected transformer is connected to the mains to supply a test current of up to 30A for "protective earth" measurements. 6.1 Theory of Operation The QA-90 is an electrical safety analyzer based on a 68HCll Motorola microprocessor and a measuring system that is galvanicly isolated from the processor. The unit is controlled from the front panel and it is possible to select either manual mode for measuring voltage, current and resistance, or an automatic test cycle that provides full safety testing in accordance with specified standards. The following standards may be compiled either fully or in part: IEC 60601.1, UL 2601.1, IEC 60601.1.1, UL 2601.1.1, IEC 60601.2.4, IEC 61010-1, EN 60601-1, VDE 0750 Tl/12-91, BS 5724, CAN/CSA-C22.2 No 601.1-M90, AS 3200.1, NZS 6150:1990, VDE 0751:10-1990, VDE 0751:10-2001, ÖVE 0751, UL 544, HEI 95, DB 9801, HEI 158 among others. The measurement results, which are shown in the unit's LCD display, may be stored in the unit's internal memory, or printed out on an external printer. A serial port (RS-232c) enables the unit to be controlled from a PC, simply by using the ansur or PRO-Soft software programs. The unit has a separate input for a bar code pen, which may be used to record instrument code and class, module code, and number of patient leads and classes. The instrument is controlled from an alphanumeric keypad, containing nine blue "soft-keys," and seven function keys located around the LCD display. These are used to select the functions shown in the display. In addition, there is a separate key for selecting patient leads, a CLR a return key and a confirmation key. All terminals for connecting the machine under test to are located on the front panel. The mains contact is located on the rear panel adjacent to the mains switch marked Power QA-90.

6-1

This page intentionally left blank.

6-2

APPENDIX A: IEC 60601.1, UL 2601.1 AND VDE 0751 TESTING
This appendix describes International Electrotechnical Committee (IEC) Standard 60601.1, Underwriters Laboratories (UL) Standard 2601.1, and Verband Deutscher Elektrotechniker e.V. (VDE) 0751 Standard tests, their functions, applicability, and equipment connections. A.1 Classification of Equipment Electrical safety begins with considering the mains electricity supply, and how to feed that into an item of equipment so that it is able to power the electronics internally and, at the same time, ensure that there is no possibility of the mains power coming into contact with either the patient, user or a third person. The classification of equipment under IEC 601.1 and UL 2601.1 describe how the mains part insulation is achieved. The techniques used include: air clearance; insulating materials (basic or functional insulation); creepage distances, and; double insulation 1. Class 0. The mains part is completely separated from any of the accessible parts. The separation is achieved by basic insulation. This is not used in medical applications.

2.

Class 1. In addition to the basic insulation, there is protection via the protective earth conductor in the mains lead. This is intended to connect all the equipment's accessible parts to earth. The majority of electromedical equipment is Class 1 equipment.

3.

Class 2. Equipment of this class is constructed with double insulation, i.e., two distinct insulating layers around the mains part. The purpose is that, should the basic insulation of the mains part fail, and then a second insulating barrier exists to prevent the mains from coming into contact with the user or patient. (See illustration below.)

A-1

4.

Class 3. Equipment of this classification is no longer manufactured, and the classification was removed from IEC 601.1 in 1988. The equipment was powered via an isolating, or safety transformer, which generated at its secondary winding a "Medical Safety Extra Low Voltage (MSELV).

5.

Symbols. A particular degree of protection afforded a patient against an electric shock, arriving from the applied part, defines the type of applied part. The symbols for each type of protection are shown below:
Symbols Description
Type B

Symbols

Description
Type B Defib Protection

Type BF

Type BF Defib Protection

Type CF

Type CF Defib Protection

B=

Classes 1, 2 or 3 equipment, or I.P. equipment providing an adequate degree of protection against electrical shock, particularly regarding allowable leakage currents and reliability of the protective earth connection.

BF = Type B equipment with an F-t