Text preview for : HP 8514B_252C 12A Test Set Operation.pdf part of Agilent HP 8514B 252C 12A Test Set Operation Agilent HP 8514B_252C 12A Test Set Operation.pdf



Back to : HP 8514B_252C 12A Test Se | Home

Errata


Title & Document Type: 8514B S-Parameter and 8512A Reflection Transmission
Test Sets Manual
Manual Part Number: 08514-90014


Revision Date: 1990-06-01



HP References in this Manual

This manual may contain references to HP or Hewlett-Packard. Please note that Hewlett-
Packard's former test and measurement, semiconductor products and chemical analysis
businesses are now part of Agilent Technologies. We have made no changes to this
manual copy. The HP XXXX referred to in this document is now the Agilent XXXX.
For example, model number HP8648A is now model number Agilent 8648A.

About this Manual

We've added this manual to the Agilent website in an effort to help you support your
product. This manual provides the best information we could find. It may be incomplete
or contain dated information, and the scan quality may not be ideal. If we find a better
copy in the future, we will add it to the Agilent website.


Support for Your Product

Agilent no longer sells or supports this product. You will find any other available
product information on the Agilent Test & Measurement website:
www.tm.agilent.com

Search for the model number of this product, and the resulting product page will guide
you to any available information. Our service centers may be able to perform calibration
if no repair parts are needed, but no other support from Agilent is available.
HP 8514B S-PARAMETER
and
HP 8512A REFLECTION/TRANSMISSION
TEST SETS
SERIAL NUMBERS

This manual applies directly to HP 8514B test sets with serial
prefix 2706A and to HP 8512A test sets with serial prefix
2631A.
For additional information about serial numbers, refer to
INSTRUMENTS COVERED BY MANUAL in the General
Information section.




0 Copyright HEWLETT-PACKARD COMPANY 1987
1400 FOUNTAINGROVE PARKWAY, SANTA ROSA, CA 95403 U.S.A.




MANUAL PART NO. 08514-90014 Printed: JUNE 1990
Microfiche Part Number 08514-90015



HEWLETT
PACKARD
CERTIFICATION
Hewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the
factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National
Institute of Standards and Technology (N/ST, formerly NBS), to the extent allowed by the Institute's calibration facility, and
to the calibration facilities of other International Standards Organization members.

WARRANTY
This Hewlett-Packard system product is warranted against defects in materials and workmanship for a
period corresponding to the individual warranty periods of its component products. Computer and computer
peripherals are warranted for a period of 90 days. Instruments are warranted for a period of one year. During
the warranty period, Hewlett-Packard Company will, at its option, either repair or replace products which
prove to be defective.
Warranty service for products installed by HP and certain other products designated by HP will be performed
at Buyer's facility at no charge within HP service travel areas. Outside HP service travel areas, warranty
service will be performed at Buyer's facility only upon HP's prior agreement and Buyer shall pay HP's round
trip travel expenses. In all other areas, products must be returned to a service facility designated by HP.
For products returned to HP for warranty service, Buyer shall prepay shipping charges to HP and HP shall
pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties,
and taxes for products returned to HP from another country.
HP warrants that its software and firmware designated by HP for use with an instrument will execute its
programming instructions when properly installed on that instrument. HP does not warrant that the opera-
tion of the instrument, or software, or firmware will be uninterrupted or error free.

LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by
Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of
the environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HP SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE REMEDIES. HP SHALL NOT
BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES,
WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.

ASSISTANCE
Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard products.
For any assistance, contact your nearest Hewlett-Packard Sales and Service Office. Addresses are provided at the back of
this manual.
8P24.2
DECLARATION OF CONFORMITY
according to ISO/IEC Guide 22 and EN 45014
Manufacturer's Name: Hewlett-Packard Co.

Manufacturer's Address: 1400 Fountaingrove Parkway
Santa Rosa, California 95403
U.S.A.

Declares that the product:

Product Name: S-Parameter Test Set

Model Numbers: HP 8514B

Product Options: This declaration covers all options
of the above product(s).

Conforms to the following product specifications:

Safety: IEC 348:1978/HD 401:1980
CAN/CSA-22.2 No. 231 Series M89
EMC: CISPR 11:1990 /EN 55011:1991, Group 1 Class A
IEC 801-2:1991 /EN 50082-1:1992,4 kV CD, 8 kV AD
IEC 801-3:1984 /EN 50082-1:1992,3 V/m, 27-500 MHz
IEC 801-4:1988 /EN 50082-1:1992, 500 V signal, 1000 V AC

Supplementary Information:

The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and
the EMC Directive 89/336/EEC.

The HP 8514B was qualified as part of a product family which includes the HP 851OC, HP 8530A,
HP 8511A, HP 8511B, HP

Santa Rosa, California

Location

European Contact:
Your local Hewlett-Packard Sales and Service Office or Hewlett-Packard GmbH, Department
ZQ/Standards Europe, Herrenberger Stra8e 130, D-7030 Biiblingen (FAX: $49-7031-143143)



Notice for Germany: Noise Declaration LpA < 70 dB
am Arbeitsplatz (operator position)
normaler Betrieb (normal position)
nach DIN 45635 T. 19 (per IS0 7779)




Regulatory Information
HP 85148/8512A Operating and Service Manual

TABLE OF CONTENTS


SECTION 1. GENERAL INFORMATION ........... 1-l HP 8512A Operator's Check . . . . . . . . . . . . . . . . . 3-6
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-l a1 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Verifying the Test Set . . . . . . . . . . . . . . . . . . . . . . . l-l b1 Reflection Test . . . . . . . . . . . . . . . . . . . . . . . 3-6
Instruments Covered by Manual . . . . . . . . . . . . . . l-2 b2 Thru Test . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Description and Operating Characteristics Controlling Multiple Test Sets . . . . . . . . . . . . . . . . 3-7
of the Instrument . . . . . . . . . . . . . . . . . . . . . . . l-2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Option 001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4 Initialization at Power-up . . . . . . . . . . . . . . . . . 3-7
Option 002 (HP 8514B only) . . . . . . . . . . . . . . . . l-4 Selecting a Test Set . . . . . . . . . . . . . . . . . . . . . 3-9
Option 003 (HP 8514B only) . . . . . . . . . . . . . . . . l-4 Measurement Calibration . . . . . . . . . . . . . . . . . 3-9
Option 002/003 (HP 85148 only) . . . . . . . . . . . . l-4 Operational Checks . . . . . . . . . . . . . . . . . . . . . 3-9
Option908 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-4 Performance Verification . . . . . . . . . . . . . . . . . 3-9
Option 910 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-4 Cable and Anti-Rotation Clamp Installation . . . . . 3-11
Option913 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-4 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Option W03, Warranty Conversion . . . . . . . . . . l-4
Option W30, Extended Service . . . . . . . . . . . . . l-4 SECTION 4. SPECIFICATIONS ................ 4-1
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5 Mechanical Specifications . . . . . . . . . . . . . . . . . . 4-1
Accessories Supplied . . . . . . . . . . . . . . . . . . . . l-5 Supplemental Characteristics . . . . . . . . . . . . . . . . 4-2
Accessories Available . . . . . . . . . . . . . . . . . . . . l-5
Operating and Safety Precautions . . . . . . . . . . . . . l-6 SECTION 5. TEST SET TROUBLESHOOTING . . . . . 5-l
Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-6 Test Set Temperature . . . . . . . . . . . . . . . . . . . . . . 5-2
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-6 Check all Connections . . . . . . . . . . . . . . . . . . . . . 5-2
Additional Equipment Required . . . . . . . . . . . . . . . l-6 Check Power Supply/Regulator, Fuses,
and Switches . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
SECTION 2. INSTALLATION . . . . .......... 2-l A15 Regulator Board Assembly . . . . . . . . . . . . 5-2
Introduction . . . . . . . . . . . . .......... 2-l Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Initial Inspection . . . . . . . . . . . .......... 2-l Address Switches . . . . . . . . . . . . . . . . . . . . . . . 5-3
Environmental Considerations .......... 2-1 Test Set Self-Test Indicators . . . . . . . . . . . . . . . . 5-3
Operation and Storage . . . .......... 2-l If Self-test Fails to Run . . . . . . . . . . . . . . . . . . . 5-3
Preparation for Use . . . . . . . . . .......... 2-l Configuration Switch . . . . . . . . . . . . . . . . . . . . . . . 5-4
Positioning the Test Set . . . . . .......... 2-l Check VTO/Driver (LO) . . . . . . . . . . . . . . . . . . . . . 5-5
Connecting the Test Set . . . . .......... 2-2 Test Set Self-Test Indicators . . . . . . . . . . . . . . . . 5-5
Packaging . . . . . . . . . . . . . . . . . . .......... 2-5 Test Set Troubleshooting Using Time Domain . . 5-6
Test Set Assembly Replacement Procedures . . . 5-7
SECTION 3. OPERATION ...... .............. 3-l (1) Bias Tee . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Introduction . . . . . . . . . . . . . . . . .............. 3-l (2) Coupler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-l 0
Front Panel Features . . . . . . . . .............. 3-2 (3) Switch/Splitter . . . . . . . . . . . . . . . . . . . . . . . . . 510
Rear Panel Features . . . . . . . . . .............. 3-3 (4) Frequency Converter . . . . . . . . . . . . . . . . . . . . 5-10
Operator's Check . . . . . . . . . . . .............. 3-4 (5) Sampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-l 1
Equipment . . . . . . . . . . . . . . .............. 3-4 (6) VT0 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Procedure . . . . . . . . . . . . . . . . . .............. 3-4 (7) Regulator Board Assembly . . . . . . . . . . . . . . . 5-l 1
HP 85148 Operator's Check . . .............. 3-4 (8) Step Attenuator Assembly . . . . . . . . . . . . . . . . 5-l 2
Sampler Test . . . . . . . . . . . . .............. 3-4 (9) Capacitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-l 2
b1 Thru Test . . . . . . . . . . . . . .............. 3-5 (10) 3.5 mm RF Connector Repair . . . . . . . . . . . . 5-12
b2 Thru Test . . . . . . . . . . . . . .............. 3-5 Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12




HP 8514B/85l2A Test Sets Table of Contents i
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-l 3 Manufacturer Codes and
(11) Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13 Reference Designations . . . . . . . . . . . . . . . . . 6-2
Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14 HP 8514B Replaceable Parts . . . . . . . . . . . . . . 6-3
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-l 4 HP 8514B Top View Showing Locations of
(12) Port 1 3.5 mm Connector Major Assemblies . . . . . . . . . . . . . . . . . . . . . . 6-7
Port 2 3.5 mm Connector . . . . . . . . . . . . . . . 5-14 HP 8514B Miscellaneous Mecnahical, Chassis,
Center Pin Repair Procedure . . . . . . . . . . . . . . 5-l 5 and Electrical Parts . . . . . . . . . . . . . . . . . . . . . 6-8
3.5 mm Nut Repair Procedure . . . . . . . . . . . . . 5-15 HP 8514B and HP 8512A Motherboard
(13) Power Transformer . . . . . . . . . . . . . . . . . . . . 5-15 Component Location Diagram . . . . . . . . . . . . 6-13
Unratioed Power Test . . . . . . . . . . . . . . . . . . . . . . 5-l 7 Parts Unique to HP 85148 and
HP 8512A Test Set . . . . . . . . . . . . . . . . . . . . . . . . 5-17 8512A Option 001 (1 of 2) . . . . . . . . . . . . . . . . 6-14
a1 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18 Parts Unique to HP 85148 Option 002 . . . . . . . 6-16
b1 Reflection Test . . . . . . . . . . . . . . . . . . . . . . . 5-l 8 HP 85148 Ports 1 and 2 Replaceable Parts . . 6-17
b2 Thru Test . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-l 8 Parts Unique to HP 85148 Option 003 . . . . . . . 6-l 8
HP 85148 Test Set . . . . . . . . . . . . . . . . . . . . . . . . 5-18 HP 8514B Option 002/003 Parts . . . . . . . . . . . 6-20
Sampler Test . . . . . . . . . . . . . . . . . . . . . . . . . . 5-l 8 HP 8512A Replaceable Parts . . . . . . . . . . . . . . 6-22
b1 Thru Test . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19 HP 8512A Top view Showing Locations of
b2 Thru Test . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-l 9 Major Assemblies . . . . . . . . . . . . . . . . . . . . . . 6-26
Next Step . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19 HP 8512A Miscellaneous Mechanical, Chassis
If Any User Channel Looks Bad . . . . . . . . . . . . . . 5-29 and Electrical Parts . . . . . . . . . . . . . . . . . . . . . 6-27
Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29 HP 8512A Port 1 and Port 2
Service Procedure . . . . . . . . . . . . . . . . . . . . . . 5-30 Replaceable Parts . . . . . . . . . . . . . . . . . . . . . 6-32
Service Adapter Conclusions . . . . . . . . . . . . . . 5-30
SECTION 7. SERVICE . . . . . . . . . . . . . . . . . . . . . . . . 7-l
SECTION 6. REPLACEABLE PART6 ............ 6-l SECTION 6. PERFORMANCE TESTS . . . . . . . . . . . . 8-l
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-l
SECTION 9. ADJUSTMENTS .................. 9-l
Exchange Assemblies Available . . . . . . . . . . . . . . 6-l
Replaceable Parts Lists . . . . . . . . . . . . . . . . . . . . . 6-2 SECTION 10. MANUAL BACKDATING . . . . . . . . . . . 1 o-1




ii Table of Contents HP 8514B/8512A Test Sets
Figure l-l. HP 8514B (upper) and HP 8512A Test Sets



1-0 General Information HP 8514B/85lZA Test Sets
Section 1. General Information



The purpose of this manual is to enable you to use your HP 8514B S-parameter or HP 8512A reflection/
transmission test set effectively and confidently. These test sets are integral components of the
HP8510 measurement system. For that reason, this manual has been divided into two major portions
(Operating and Service) to be an integral part of the HP8510 documentation.
l General Information
0 Installation
0 Operation
l Performance Tests
l Adjustments
l Backdating


The Service part consists of :
l Replaceable Parts
0 Service
The major topics of this section, General information, are:
l how to use the test set
l what the test set is
l operating, safety and warranty considerations
0 test set specifications



VERIFYING THE TEST SET

The HP8514B and HP8512A have been designed to operate specifically with the HP8510 network
analyzer.
0 To install the instrument, turn to the Installation section of this manual.
l To check the proper operation of the test set, see the Operator's Check in the Operation section of
this manual.

l To see the specifications of the test set refer to Specifications in the HP8570B System Manual or
Genera/ information in volume 1 of the HP 8510A manual set.




HP 8514B/85l2A Test Sets General Information l-l
l To verify that the instrument meets its published specifications, turn to the Performance Tests
section in the HP8510B System Manual or Volume 2 of the HP851OA Operating and Service Manual.
Note that the HP8514B can be performance tested only to 18 GHz with the HP851OA software.
l To troubleshoot the test set, refer to the Service Overview section and the TestSet Troubleshooting
section of the HP8510B Service Manual. Or refer to the Service section in Volume 4 of the HP851OA
Operating and Service Manual. Otherwise call your local Hewlett-Packard office.



INSTRUMENTS COVERED BY MANUAL

You will find a two-part serial number on the rear panel of the instrument. The first four digits and the
letter are the serial number prefix. The last five digits are the sequential suffix which is unique to each
test set. The contents of this manual apply directly to test sets with the same serial number prefix as the
one(s) on the title page under the heading Serial Numbers.
If the serial prefix of your test set is not listed on the title page, your instrument differs from those
documented in this manual. The differences are documented in the yellow manual changes supplement
supplied with the manual.
To keep this manual as current and accurate as possible, Hewlett-Packard recommends that you
periodically request the latest manual changes supplement, as it may contain replacement information
as well as change information. The supplement for this manual is keyed to the manual's print date and
part number (on the title page) and is available on request from Hewlett-Packard.
You can order this manual in microfiche form (the part number appears on the title page). With the
manual (in 4 x 6 inch microfilm transparency format) you will also receive the latest manual changes
supplement.



DESCRIPTION AND OPERATING CHARACTERISTICS
OF THE INSTRUMENT

The combination of the HP 8514B test set with the HP8510 network analyzer and source provides a
system for making S-parameter measurements over the frequency range of 45 MHz to 20 GHz. This
system is suited for making measurements on two port devices when it is inconvenient or inexpedient
to physically reverse the DUT (device under test) to measure all four S-parameters.
The HP8514B uses two couplers for signal separation. For measurements of active devices, the
standard HP 8514B includes:
l four RF to IF converters to measure all four S-parameters without reconnecting the DUT,
l two 90 dB programmable step attenuators for changing (in 10 dB steps) the incident power
level at both ports,
l two bias tees for applying external dc bias to both test port center conductors.




1-2 General Information HP 8514B/85l2A Test Sets
In the "High Forward Dynamic Range Configuration" (option 003) the Port 2 coupler is reversed to
optimize dynamic range in the forward measurement direction. Because the b2 sampler is connected
to the coupler through path instead of the coupled arm, there is less isolation between the b2 sampler
and Port 2, and the b2 power level will be higher than bl. This configuration is better for wide dynamic
range or reciprocal devices or components like filters, cables, or antennas, where S,, is generally not
measured. However, as a result of the lower isolation between the test port and the sampler, "sampler
bounce," appearing as an occasional spurious response related to the VT0 frequency, can be
noticeable in this "asymmetrical" test set.
The HP 8512A reflection/transmission test set uses one coupler for signal separation. The standard
HP 8512A has:
l three RF to IF converters, for making reflection (S,,) and transmission (S,,) measurements,
l no step attenuators to internally change the incident power level,
l no bias tees to apply external DC bias to the test port center conductors.
Tables 1 and 1B list additional characteristics of the HP85148 and 8512A, respectively. Figure l-2
shows the HP 8514B in a typical measurement set-up.


8510 INTERCONNECT

~ 1 f ,J15 SOURCE CONTROL 1 ~ ! (




RF OUTPUT
RF INPUT T E S T S E T




TEST SET H P 8510B HP 83YOO R 83'41
NETYORK ANALYZER SYNTHESIZE0
WEEPER


Figure 1-2. Typical HP8514B Measurement Set-up




HP 8514B/8512A Test Sets General Information 1-3
Option 001
This option adds IF switching capability to allow up to four test sets to be connected to the HP8510B at
the same time. The test set in use is selected from the HP85108 front panel. The 20 MHz IF signal is
transmitted from the standard test set through the option 001 test set(s) to the network analyzer. IF
switching is performed automatically by the option 001 test set(s), without reconnections.

Option 002 (HP 85148 only)
This option deletes the 90 dB programmable step attenuators and the dc bias tees. Note that bias can
be applied externally, using the HP11612A bias tee, if bias is required but attenuation not.

Option 003 (HP 85148 only)
This high forward dynamic range configuration is described in detail in this section under the title
Description and Operating Characteristics of the Instrument.

Option 002/003 (HP 85148 only)
This option is a combination of option 002 and option 003, which were previously described.

Option 908
This option supplies the test set with the parts required to rack mount it with handles removed. Refer to
the installation section of this manual for additional information.

Option 910
This option provides a duplicate test set manual.

Option 913
This option supplies the test set with the parts required to rack mount it with handles. Refer to the
INSTALLATION section of this manual for additional information.

Option W03, Warranty Conversion
Option W03 converts the standard one year return to Hewlett-Packard warranty to a 90 day on-site
warranty. W03 can only be ordered at the time of instrument purchase. Instruments ordered with option
W03 are identified on the serial number label, or on a special identification label supplied with the
instrument.

Option W30, Extended Service
Option W30 adds two additional years of return-to-HP service, to follow the first year of warranty.
Option W30 can be ordered only at the time of purchase. Instruments ordered with option W30 are
identified on the serial number label, or on a special identification label supplied with the instrument.
NOTE: additional system warranty information is included in the HP8510 manual set.




1-4 General Information HP 8514B/85l2A Test Sets
ACCESSORIES

Accessories Supplied
Figure 3-1 shows the accessories supplied with the HP8514B and 8512A test sets (except as noted in
Table 3-l). The accessories, with part numbers, are listed in the installation section for both test sets.

Accessories Available
NOTE: Additional HP8510 system accessory information is located in the HP8510 manual set.
Calibration, Verification and Adapter Kits. Hewlett-Packard offers several calibration kits suitable
for calibrating an HP8510/8514B or 8512A when making error corrected measurements. Each calibra-
tion kit noted below includes a set of precision standards to calibrate an HP8510 system in the indicated
interface. Additional information is located in the System and Documentation Overview section of the
HP8570B System Manual and volume 1 of the HP8510A manual set.


Connector Type Calibration Kit Verification Kit Adapter Kit
3.5 mm HP850520/E HP85053B
7mm HP 850508/C/D HP 85051 B
Type-N 5OQ HP 85054B HP 85055A
3.5 to 7 mm HP 85130A/B'

1. These adapters attach directly to the 3.5 mm test ports.



RF Cables. The HP 85131D 3.5 mm Test Port Return Cable Set is a pair of 21 inch long cables specified
from DC to 26.5 GHz. Typically it is used with the HP8514B. One of the cables has 3.5 mm (f)
connectors, the other cable has one 3.5 mm (f) and one 3.5 mm (m) connector.
The HP85132C 7 mm Test Port Return Cable is a single, 36 inch cable for measurements where the
device is connected directly to one test port. It is typically used with the HP8512A. The cable has two
7 mm connectors. Its frequency range is DC to 18 GHz.
Transistor Test Fixture Kit. The HP85041A Transistor Test Fixture Kit (TTF) is a comprehensive
measurement system for testing and characterizing stripline packaged microwave transistors.
Although it has 7 mm connectors and a frequency range limited to 18 GHz, the TTF may be easily
adapted for use with the HP 85148. Please consult with your local HP Systems Engineer for specific
recommendations.




HP 8514B/85l!2A Test Sets General Information l-5
Operating




AITENTION
Static Sensitive
Handle only at Static Safe
Work Stations

Beware of electro-static damage (ESD). The input connectors (test ports
or cables or adapters connected to the test ports) are very sensitive to
ESD. Use a grounded wrist strap when attaching devices to the input
connectors.
Otherwise, you need observe only normal precautions in handling and operating the test set. Do not
exceed the front panel operating level power input as noted:
Maximum Operating Power Level Test Port
f17 dBm HP8514B Port 1 and 2
+20 dBm HP 8512A Port 1
-13 dBm HP8512A Port 2
l Do not torque anything to the test port connector with greater than 90 N-cm (8in.-lb.) of torque. The
wrench supplied with your accessory kit is calibrated 90 N-cm (8 in.-lb.).
l Do not torque anything to the Source RF input, on the back of your test set, with greater than 90 N-
cm (8 in.-lb.).

Service
The voltages in this test set warrant normal caution for operator safety. Nevertheless, service should
be performed only by qualified personnel. Service strategy, troubleshooting procedures, replaceable
parts and similar information for the HP 85148 and 8512A test sets is in the HP85708 Service Manual.




Table 2-5 lists additional equipment and accessories required for use with the HP 8514B and 8512A test
sets. The table notes which items are required to verify the performance of the test sets and which are
required to operate them. Other equipment may be substituted if its specifications meet or exceed the
specifications listed in the critical specifications column.




l-6 General Information HP 8514B/8512A Test Sets
Section 2. Installation

INTRODUCTION

This section explains how to install the HP 8514B and HP8512A test sets. The topics covered include
initial inspection, environmental considerations, positioning and connecting the test set for use, and
packaging the instrument. Refer to the Installation section of the HP8510 manual for more complete
system connection and turn-on instructions.




Inspect the shipping container (including cushioning material) for damage. If it is damaged, keep it until
you have checked the contents for completeness. The contents are listed and illustrated in Figure 2-l.
In addition, check the test set mechanically and electrically. If the test set and shipping container are
undamaged, passing the Operator's Check (in the Operation section) should suffice for incoming
inspection. If the test set does not pass the Operator's Check, refer to the troubleshooting procedures
in the service portion of this manual. Alternatively, call your local HP customer engineer.
If the shipping container is damaged, perform the performance tests outlined in the HP8510 manual
set. If the test set fails the performance tests, or is damaged or defective, keep the shipping materials
and notify both the carrier and the nearest Hewlett-Packard office. The HP office will arrange for repair
or replacement of the test set without waiting for settlement of the claim. If any of the following
accessories are not received with the test set, notify your nearest HP office and the missing parts will be
sent to you.




Operation and Storage
To perform within specifications, the test sets should be operated in temperatures between 0