Agilent 42030A Four-Terminal Pair Standard
Resistor Set
Operation and Service
Manual




Agilent Part No. 42030-90001
Printed in JAPAN January 2000
Second Edition
CERTIFICATION Agilent Technologies certi es that this product met its published
speci cations at the time of shipment from the factory. Agilent
Technologies further certi es that its calibration measurements are
traceable to the United States National Institute of Standards and
Technology (nist), to the extent allowed by the Institute's calibration
facility, or to the calibration facilities of other International Standards
Organization members.

WARRANTY This Agilent Technologies instrument product is warranted against
defects in material and workmanship for a period of one year
from the date of shipment. During the warranty period, Agilent
Technologies will, at its option, either repair or replace products
which prove to be defective.
For warranty service or repair, this product must be returned to a
service facility designated by Agilent Technologies. Buyer shall prepay
shipping charges to Agilent Technologies and Agilent Technologies
shall pay shipping charges to return the product to Buyer. However,
Buyer shall pay all shipping charges, duties, and taxes for products
returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and rmware
designated by Agilent Technologies for use with an instrument will
execute its programming instruction when properly installed on that
instrument. Agilent Technologies does not warrant that the operation
of the instrument, or software, or rmware will be uninterrupted or
error free.

LIMITATION OF The foregoing warranty shall not apply to defects resulting from
WARRANTY improper or inadequate maintenance by Buyer, Buyer-supplied
software or interfacing, unauthorized modi cation or misuse,
operation outside of the environmental speci cations for the product,
or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT
TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.

EXCLUSIVE THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND
REMEDIES EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE
LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT,
TORT, OR ANY OTHER LEGAL THEORY.

ASSISTANCE Product maintenance agreements and other customer assistance
agreements are available for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales
and Service Oce. Addresses are provided at the back of this manual.
Notice The information contained in this document is subject to change
without notice.
This document contains proprietary information which is protected by
copyright. All rights are reserved. No part of this document may be
photocopied, reproduced, or translated to another language without
the prior written consent of the Agilent Technologies.

Agilent Technologies Japan, Ltd.
Component Test PGU-Kobe
1-3-2, Murotani, Nishi-ku, Kobe-shi,
Hyogo, 651-2241 Japan




Copyright 1992,2000 Agilent Technologies Japan, Ltd.
c
Manual Printing The manual printing date and part number indicate its current
History edition. The printing date changes when a new edition is printed.
(Minor corrections and updates which are incorporated at reprint do
not cause the date to change.) The manual part number changes when
extensive technical changes are incorporated.
October 1992 : : : : : : : : : : : : : : First Edition (part number: 42030-90001)
January 2000 : : : : : : : : : : : Second Edition (part number: 42030-90001)




iv
How To Use This Manual. This is the Operation and Service Manual
for the 42030A Four-Terminal Pair Standard Resistor Set. This
contains operation and service information, and consists of the
following two chapters.

Chapter 1 General Chapter 1 describes the 42030A description, speci cation, and, other
Information general information.

Chapter 2 Service Chapter 2 describes the 42030A calibration and repair.




v
Contents
1. General Information
Introduction . . . . . . . . . . . . . . . . . . . . . 1-1
Description . . . . . . . . . . . . . . . . . . . . . . 1-1
Initial Inspection . . . . . . . . . . . . . . . . . . . 1-2
Repackaging for Shipment . . . . . . . . . . . . . . 1-2
Agilent Packing Materials . . . . . . . . . . . . . . 1-2
Non-Agilent Packing Materials . . . . . . . . . . . . 1-2
Speci cations . . . . . . . . . . . . . . . . . . . . 1-3
DC Resistance . . . . . . . . . . . . . . . . . . . 1-3
Operating Conditions . . . . . . . . . . . . . . . . 1-3
Dimensions . . . . . . . . . . . . . . . . . . . . . 1-3
Weight . . . . . . . . . . . . . . . . . . . . . . . 1-4
Supplemental Performance Characteristics . . . . . . . 1-4
Temperature Coecient and Stability . . . . . . . . 1-4
Storage Conditions . . . . . . . . . . . . . . . . . 1-4
2. Calibration and Repair
Introduction . . . . . . . . . . . . . . . . . . . . . 2-1
Calibration . . . . . . . . . . . . . . . . . . . . . . 2-1
Calibration Interval . . . . . . . . . . . . . . . . . 2-4
Repair . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Index




Contents-1
Tables
1-1. 42030A Contents . . . . . . . . . . . . . . . . . . 1-1
1-2. DC Resistance . . . . . . . . . . . . . . . . . . . 1-3
1-3. Operating Conditions . . . . . . . . . . . . . . . . 1-3
1-4. Dimensions . . . . . . . . . . . . . . . . . . . . . 1-3
1-5. Weight . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-6. DC Resistance Temperature Coecient and Stability . 1-4
1-7. Storage Conditions . . . . . . . . . . . . . . . . . 1-4
2-1. 42030A Calibration Parameters . . . . . . . . . . . 2-1
2-2. 42030A DC Resistance Calibration Uncertainty . . . . 2-2
2-3. 42035A 10
Resistor AC Calibration Uncertainty . . . 2-2
2-4. 42036A 100
Resistor AC Calibration Uncertainty . . 2-3
2-5. 42037A 1 k
Resistor AC Calibration Uncertainty . . . 2-3
2-6. 42038A 10 k
Resistor AC Calibration Uncertainty . . 2-3
2-7. 42039A 100 k
Resistor AC Calibration Uncertainty . . 2-4




Contents-2
1
General Information
Introduction This chapter provides the 42030A Four-Terminal Pair Standard
Resistor Set description, speci cations, and related general
information.




Description The 42030A Four-Terminal Pair Standard Resistor Set consists of
nine precision resistor standards. The set is constructed using high
stability resistors and are virtually unaected by changes in ambient
temperature, ensuring measurement repeatability and reliability.
Table 1-1 lists the 42030A contents.
Table 1-1. 42030A Contents
Model No. or Description Qty.
Agilent Part No.
42031A 1 m
Standard Resistor 1
42032A 10 m
Standard Resistor 1
42033A 100 m
Standard Resistor 1
42034A 1
Standard Resistor 1
42035A 10
Standard Resistor 1
42036A 100
Standard Resistor 1
42037A 1 k
Standard Resistor 1
42038A 10 k
Standard Resistor 1
42039A 100 k
Standard Resistor 1
42030-60100 Carrying Case 1
42030-90001 Operation and Service Manual 1
Not Assigned Calibration Report 1

The 42030A is designed for use in calibrating the Agilent LCR meters
which have four-terminal pair measurement terminals, and can be
directly connected to the measurement terminals. The lcr meter
documentation will give detailed information on using the 42030A.




General Information 1-1
Initial Inspection Inspect the shipping container for damage. If the shipping container
or cushioning material is damaged, it should be kept until the shipping
contents have been checked for completeness and the instrument has
been checked mechanically and electrically. The shipping contents
should consist of the items shown in Table 1-1. If the shipment is
incomplete, or if there is mechanical damage or other defects, notify
your nearest Agilent Technologies oce. If the shipping container is
damaged, or the cushioning material shows signs of stress, notify the
carrier as well as the Agilent Technologies oce. Keep the shipping
materials for the carrier's inspection. Your Agilent Technologies oce
will arrange for repair or replacement, without waiting for the claim
settlement.




Repackaging for
Shipment
Agilent Packing Containers and materials identical to those used in factory packaging
Materials are available from Agilent Technologies. If the instrument is being
returned to Agilent Technologies for servicing, attach a tag indicating
the type of service required, return address, model number, and full
serial number. Also, mark the container FRAGILE to ensure careful
handling. In any correspondence, refer to the unit by model number
and full serial number.

Non-Agilent Packing The following general instructions should be used for re-packing with
Materials commercially available packing materials:
a. If shipping to Agilent Technologies oce or service center, attach a
tag indicating the type of service required, return address, model
number, and full serial number.
b. Use a strong shipping container. A double-walled carton made of
350 pound test material is adequate.
c. Use enough shock absorbing material (3- to 4-inch layer) around
all sides of the case to provide a rm cushion and to prevent
movement inside the container.
d. Seal the shipping container securely.
e. Mark the shipping container FRAGILE to ensure careful handling.
f. In any correspondence, refer to the unit by model number and full
serial number.




1-2 General Information
Speci cations This section gives the complete speci cations for the 42030A
Four-Terminal Pair Standard Resistor Set. When the 42030A is shipped
from the factory, it meets the speci cations listed in this section.

DC Resistance Table 1-2 lists the nominal DC resistance for each resistance module of
the 42030A.
Table 1-2. DC Resistance
Model Number DC Resistance1
42031A 1 m
6 0.2%
42032A 10 m
6 0.2%
42033A 100 m
6 0.2%
42034A 1
6 0.2%
42035A 10
6 0.1%
42036A 100
6 0.1%
42037A 1 k
6 0.1%
42038A 10 k
6 0.1%
42039A 100 k
6 0.1%
1 Speci ed under ambient environmental
conditions of: Temperature
23 C 6 5 C , Relative Humidity
70%

Operating Conditions The 42030A must be operated under the ambient environmental
conditions listed in Table 1-3
Table 1-3. Operating Conditions
Temperature 23 C 6 5 C
Relative Humidity 70%

Dimensions Table 1-4 shows the 42030A dimensions.
Table 1-4. Dimensions
Resistance Module 94 mm (W) 2 31 mm (H) 2 67 mm (D)
Carrying Case 400 mm (W) 2 120 mm (H) 2 290 mm (D)




General Information 1-3
Weight Table 1-5 shows the 42030A weight.
Table 1-5. Weight
Resistance Module 170 g (ea.)
42030A Total 3,700 g




Supplemental This section gives the supplemental performance characteristics
Performance for the 42030A Four-Terminal Pair Standard Resistor Set . These
supplemental performance characteristics are not guaranteed, they
Characteristics are the 42030A typical characteristics that may be expected.

Temperature Table 1-6 shows the 42030A DC resistance temperature coecient and
Coecient and stability.
Stability Table 1-6.
DC Resistance Temperature Coecient and Stability
Model Number Nominal Temperature Stability2
Value Coecient1
42031A 1 m
6 15 ppm/ C 6100 ppm/year
42032A 10 m
6 15 ppm/ C 6100 ppm/year
42033A 100 m
6 15 ppm/ C 6100 ppm/year
42034A 1
6 15 ppm/ C 6100 ppm/year
42035A 10
6 10 ppm/ C 6100 ppm/year
42036A 100
6 10 ppm/ C 6100 ppm/year
42037A 1 k
6 10 ppm/ C 6100 ppm/year
42038A 10 k
6 10 ppm/ C 6100 ppm/year
42039A 100 k
6 10 ppm/ C 6100 ppm/year
1 Under temperature condition: 23 C 6 5 C
2 Under environmental conditions: Temperature 23 C 6 5 C , Relative
Humidity 70%


Storage Conditions The 42030A must be stored or shipped under the ambient
environmental conditions listed in Table 1-7.
Table 1-7. Storage Conditions
Temperature 0 40 C to + 70 C
Relative Humidity 95% at 40 C



1-4 General Information
2
Calibration and Repair
Introduction This chapter provides the calibration and repair information for the
42030A Four-Terminal Pair Standard Resistor Set .




Calibration The 42030A is calibrated in the parameters given in Table 2-1 with
the calibration uncertainties given in Table 2-2 through Table 2-7 and
equipped with a calibration report, when shipped from the factory.
Table 2-1. 42030A Calibration Parameters
Model Number Nominal DC Resistance Series Resistance Parallel Resistance
Value and Reactance and Susceptance
42031A 1 m

p
42032A 10 m

p
42033A 100 m

p
42034A 1

p
42035A 10

p p
42036A 100

p p
42037A 1 k

p p
42038A 10 k

p p
42039A 100 k

p p




Calibration and Repair 2-1
Table 2-2.
42030A DC Resistance Calibration Uncertainty
Model Number Nominal Calibration
Value Uncertainty
42031A 1 m
6 0.1%
42032A 10 m
6 0.1%
42033A 100 m
6 0.1%
42034A 1
6 0.1%
42035A 10
6 0.03%
42036A 100
6 0.02%
42037A 1 k
6 0.02%
42038A 10 k
6 0.02%
42039A 100 k
6 0.02%

Table 2-3.
42035A 10
Resistor AC Calibration Uncertainty
Calibration Uncertainty
Frequency
Series Resistance Reactance
1 MHz 60.12% 60.010

2 MHz 60.15% 60.010

3 MHz 60.20% 60.015

4 MHz 60.25% 60.025

5 MHz 60.4% 60.03

10 MHz 61.0% 60.08

13 MHz 61.5% 60.10





2-2 Calibration and Repair
Table 2-4.
42036A 100
Resistor AC Calibration Uncertainty
Calibration Uncertainty
Frequency
Series Resistance Reactance
1 MHz 60.12% 60.10

2 MHz 60.12% 60.10

3 MHz 60.15% 60.15

4 MHz 60.15% 60.15

5 MHz 60.20% 60.20

10 MHz 60.5% 60.5

13 MHz 60.7% 60.7


Table 2-5.
42037A 1 k
Resistor AC Calibration Uncertainty
Calibration Uncertainty
Frequency
Parallel Susceptance
Resistance
100 kHz 60.12% 60.8 S
1 MHz 60.08% 61.0 S
2 MHz 60.10% 61.0 S
3 MHz 60.10% 61.5 S
4 MHz 60.15% 62.0 S
5 MHz 60.15% 63.0 S
10 MHz 60.4% 65 S
13 MHz 60.6% 68 S

Table 2-6.
42038A 10 k
Resistor AC Calibration Uncertainty
Calibration Uncertainty
Frequency
Parallel Susceptance
Resistance
100 kHz 60.08% 60.10 S
1 MHz 60.08% 60.20 S




Calibration and Repair 2-3
Table 2-7.
42039A 100 k
Resistor AC Calibration Uncertainty
Calibration Uncertainty
Frequency
Parallel Susceptance
Resistance
100 kHz 60.08% 60.02 S
1 MHz 60.12% 60.15 S

The 42030A must be calibrated at a facility which satis es the
calibration uncertainties given in Table 2-2 through Table 2-7. Agilent
Technologies's calibration laboratories satisfy all of these calibration
uncertainties. For complete information on calibration, contact your
nearest Agilent Technologies sales and service oce.

Calibration Interval The 42030A's recommended calibration interval is one year.




Repair The 42030A can be repaired only at Agilent Technologies facilities.
For complete information on repair, contact your nearest Agilent
Technologies Sales and Service Oce.




2-4 Calibration and Repair