Text preview for : AN150-7 Spectrum Analysis - Signal Enhancement June75.pdf part of Agilent AN150-7 Spectrum Analysis - Signal Enhancement June75 Agilent AN150-7 Spectrum Analysis - Signal Enhancement June75.pdf



Back to : AN150-7 Spectrum Analysis | Home

Errata


Title & Document Type: Application Note 150-4: Spectrum
Analysis Signal Enhancement

Manual Part Number: unknown


Revision Date: June 1975



HP References in this Manual

This manual may contain references to HP or Hewlett-Packard. Please note that Hewlett-
Packard's former test and measurement, semiconductor products and chemical analysis
businesses are now part of Agilent Technologies. We have made no changes to this
manual copy. The HP XXXX referred to in this document is now the Agilent XXXX.
For example, model number HP8648A is now model number Agilent 8648A.

About this Manual

We've added this manual to the Agilent website in an effort to help you support your
product. This manual provides the best information we could find. It may be incomplete
or contain dated information, and the scan quality may not be ideal. If we find a better
copy in the future, we will add it to the Agilent website.


Support for Your Product

Agilent no longer sells or supports this product. You will find any other available
product information on the Agilent Test & Measurement website:
www.tm.agilent.com

Search for the model number of this product, and the resulting product page will guide
you to any available information. Our service centers may be able to perform calibration
if no repair parts are needed, but no other support from Agilent is available.
" "




rj3'HEWLETT
~~ PACKARD
"
r~ ..'

CHAPTER 1
,"}
. .


Basic Considerations

Definition of Sensitivity
Sensitivity, to be useful, needs to relate to how small a signal can be measured on the analyzer. The
CRTdeflection is always proportional to the total power which includes the signal and the noise. A signal
can be seen when it is equal to the noise power.
S+N
S=Nor-=2 (1), where
N
S = power of the signal
N = power of the noise

In this case, S + N will be twice the noise power or deflected 3 dB above the noise.

Available Noise Power
The input termination of a network (an amplifier, receiver, or spectrum analyzer) has a certain amount
of available noise power which is, in most cases, thermal noise. An impedance Z = R + jX at temperature
T generates across its open circuit terminals a voltage resulting from the random motion of free electrons
thermally agitated. This "noise voltage", en, can be defined by the equation:

en: = 4kTBR,
where
k = Boltzmann's constant 1.374x 10-23joule/oK
T = absolute temperature oK
R = resistive component of impedance
B = Bandwidth
J- ",
~
If the impedance Z = R. + jX is connected to a matched load with input impedance Z = Z. as shown in
Figure 1, maximum transfer of the noise power will occur. Noise power Pn will be dissipated in the load
resistance RL due to the noise voJtage generated in the original resistance R. The noise power will be:

P = (en/2)2 =~ = 4KTBR
n RL 4RL 4RL

Since there is equal noise voltage across source and load when R = RL

Pn = KTB (2).



r '
I I
I
I
I
I
I
R+jX RL -jX I
I
I
I
I
I
I I
L ----- NETWORK - - - - .J


Filure 1. Availablenoise power P. is equal to KTB.
_\"
.~ y
1
.-
Equation (2) defines the available noise power from the source. In systems operating at frequencies
where voltages and resistances cannot be clearly defined, this equation becomes the usable expression,
containing terms that can be measured.

Noise Figure
Let us consider the network in Figure 2 with a power gain G which can be more or less than 1. In prac-
tice a network is never noiseless and decreases the signal-to-noise ratio.




St ~
G
N, N2




Fillure2.
.~
The noise figure of the network may be defined as the ratio of input signal-to-noise power ratio to the
output signal-to-noise power ratio. .

. StiNt St N2
NOIse figure F =-=-'-,where
~/N2 Nt 52
St = input signal power
Nt = input noise power
~ = output signal power
N: = output noise power
'N
Since ~ = StG F = Nt~

. H the network is noiseless, the output noise will just be equal to the amplified input noise. In other-
words, N2 = NtG and F= 1. When F > I, there is degradation of the input signal-to-noise ratio. The out-
put noise power Nt, from a noisy network is made up of two terms:
.
. The first due to the amplification of the input noise power N.G
The second is the amount of noise power generated by the noisy network and is
-
equal to (F 1) N,G. So that Nt = N.G + (F - 1) N.G = FN.G
We have seen Nt is the input noise power or the available noise powj!r; that is, Nt = KTB. It follows
then that

Nt = FkTBG (3).

Sensitivity of a Spectrum Analyzer
We can use equation (3) to figure out the output noise power or sensitivity of a spectrum analyzer.
Unfortunately, the gain is unknown and we prefer to define the total input noise power which is the
output noise power divided by the gain.
.



2
~ Equation (3) becomes

N = FsAkTB (4) with FSA = Spectrum analyzer noise figure.


x;:.' It's more convenient to express the formula in dB

10 log N = 10 log FSA+ 10 log kT + 10 log B


At room temperature, T = 290okand 10log KT= -204 dB, a value which is constant in normal utilization.
(VVehave an error of 0.4 dB for t = 23