Text preview for : 8753E_User.pdf part of Agilent 8753E User Agilent 8753E 8753E_User.pdf



Back to : 8753E_User.pdf | Home

User's Guide

HP 8753E Network Analyzer




HP Rut No. 08753-90367 Supersedes October 1998
Printed iu USA February 1999
Notice.
The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including
but not limited to, the implied warranties of merchantability and fitness for a particular
purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this material.




@ Copyright Hewlett-Packard Company 1998, 1999
Certification
Hewlett-Packard Company certifies that this product met its published specifications at the
time of shipment from the factory. Hewlett-Packard further certifies that its calibration
measurements are traceable to the United States National Institute of Standards and
`lbchnology, to the extent allowed by the Institute's calibration facility, and to the calibration
facilities of other International Standards Organization members.




Wmanty
Note The actual warranty on your instrument depends on the date it was ordered
as well as whether or not any warranty options were purchased at that time.
lb determine the exact warranty on your insment, contact the nearest
Hewlett-Packard sales or service office with the model and serial number of
your instrument. See the table titled "Hewlett-Packard Sales and Service
Offices," later in this section, for a list of sales and service offices.

This Hewlett-Packard instrument product is warranted against defects in material and
workmanship for the warranty period. During the warranty period, Hewlett-Packard Company
will, at its option, either repair or replace products which prove to be defective.
If the warranty covers repair or service to be performed at Buyer's facility, then the service or
repair will be performed at the Buyer's facility at no charge within HP service travel areas
Outside HP service travel areas, warranty service will be performed at Buyer's facility only
upon HP's prior agreement, and Buyer shall pay HP's round-trip travel expenses. In all other
areas, products must be returned to a service facility designated by HP
If the product is to be returned to Hewlett-Packard for service or repair, it must be returned
to a service facility designated by Hewlett-Packard. Buyer shall prepay shipping charges to
Hewlett-Packard and Hewlett-Packard shall pay shipping charges to return the product to
Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned
to Hewlett-Packard from another country.
Hewlett-Packard warrants that its software and lirmware designated by Hewlett-Packard for
use with an instrument will execute its programming instructions when properly installed on
that instrument. Hewlett-Packard does not warrant that the operation of the instrument, or
software, or firmware will be uninterrupted or error-free. LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or
misuse, operation outside of the environmental specifications for the product, or improper
site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY
DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE REMEDIES.
HEWLETT-PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT,
OR ANY OTHER LEGAL THEORY.


...
III
Maintenance
Clean the cabinet, using a damp cloth only.




Assistance
Product maintenance agreements and other customer assktmm agremnmts are available for
Hewlett-&hrd products
Rw an@ assistance, wnmct gour nearest Hewlett-Rzchmd Saks and Service Om




Shipment for Service
If you are sending the instrument to Hewlett-Packard for service, ship the analyzer to the
nearest HP service center for repair, including a description of any failed test and any error
message. Ship the analyzer, using the original or comparable anti-static packaging materials.




iv
`Iktble O-1. Hewlett-Packard Sales and Service OfEces
UNITED STATES

Instrument Support Center
Hewlett-Packard Company
(800) 403-0801

EUROPEAN FIELD OPEEA!l'IONS

Headquarters Prame G-m=w
Hewlett-Packard S.A. Hewlett-Packard France Hewlett-Packard GmbH
150, Route du Nantd'Avril 1 Avenue Du Canada Hewlett-Packard Strasse
1217 Meyrin Z/Geneva Zone D'Activite De Courtaboeuf 61352 Bad Homburg v.d.H
&vi&&and F-91947 Les Ulis Cedex Germany
(41 22) 780.8111 France (49 6172)16-O
(33 1) 69 82 60 60
Great Britah
Hewlett-Packard Ltd.
E&dale Road, Whmersh Triangle
Woldngham, Berkshire RG415DZ
England
(44 734) 696622

INTERCON FIELD OPERATIONS

HWkpWtt%S
Hewlett-Packard Company Hewlett-Packard Australia Ltd. Hewlett-Packard (Canada) Ltd.
3495 Deer Creek Road 31-41 Joseph Street 17500 South Service Road
Palo Alto, California, USA Blackbum, Victoria 3130 TramCanada Highway
94304-1316 (61 3) 895-2895 Kirkland, Quebec HQJ 2X8
(415) 857-5027
z7-4232

JspaJp fb@me
China Hewlett-Packard Company Hewlett-Packard Japan, Ltd. Hewlett-Packard Singapore (Pte.) Ltd.
38BeiSanHuanXlRoad Q-l %kakuraCho, Hachioji 150 Beach Road
aulaug Yu shu lblcyo 192, Japan #29-00 Gateway West
Hai Dian District (81 426) 60-2111 Singapore 0718
Beijin& china (65) 291-9088
(86 1) 256-6888

IkbiWSn
Hewlett-Packard `&wan
8th Floor, H-P Building
337 Fu I-king North Road
Tkdpei, Taiwan
(886 2) 712-0404 .




V
Safety Symbols
The following safety symbols are used throughout this manual. Familiarize yourself with each
of the symbols and its meaning before operating this instrument.

Caution Caution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, would result in damage to or destruction
of the instrument. Do not proceed beyond a caution note until the indicated
conditions are fully understood and met.

Warning Rkning denotes a hazard. It calls attention to a procedure which, if not
correctly performed or adhered to, could result in injury or loss of life.
Do not proceed beyond a warning note until the indicated conditions are
fully understood and met.




Instrument Markings
!
A The instruction documentation symbol. The product is marked with this symbol when it
is necessary for the user to refer to the instructions in the documentation.
"CE" The CE mark is a registered trademark of the European Community. (If accompanied by
a year, it is when the design was proven.)
"ISMl-A" This is a symbol of an Industrial Scientific and Medical Group 1 Class A product.
"CSA" The CSA mark is a registered trademark of the Canadian Standards Association.




vi
General Safety Considerations
Note This instrument has been designed and tested in accordance with IEC
Publication 1010, Safety Requirements for Electronics Measuring Apparatus,
and has been supplied in a safe condition. This instruction documentation
contains information and warnings which must be followed by the user to
ensure safe operation and to maintain the instrument in a safe condition.

Warning This is a safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact. Any
interruption of the protective conductor, inside or outside the instrument,
is likely to make the instrument dangerous. Intentional interruption is
prohibited.

Warning No operator serviceable parts inside. Refer servicing to qualified
personnel. lb prevent electrical shock, do not remove covers.


Caution Before switching on this instrument, make sure that the line voltage selector
switch is set to the voltage of the power supply and the correct fuse is
installed. Assure the supply voltage is in the specified range.

Warning The opening of covers or removal of parts is likely to expose dangerous
voltages. DiSCOMeCt the instrument from all voltage sources while it is
being opened.

Warning The power cord is connected tointernal capacitors that may remain live
for 10 seconds after disconnecting the plug from its power supply.


Warning For continued protection against fire hazard replace line fuse only with
same type and rating (F 3A/250V). The use of other fuses or material is
prohibited.

Warning lb prevent electrical shock, disconnect the HP 87533 from mains before
cleaning. Use a dry cloth or one slightly dampened with water to clean
the external case parts. Do not attempt to clean internally.

Warning If this product is not used as specitled, the protection provided by the
equipment could be impaired. This product must be used in a normal
condition (in which all means for protection are intact) only.

Warning Always use the three-prong AC power cord supplied with this product.
FMlure to ensure adequate earth grounding by not using this cord may
cause product damage.
Caution This product is designed for use in Installation Category II and Pollution Degree
2 per IEC 1010 and 664 respectively.


Caution VENTILATION REQUIREMENTS: When instaIling the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4O C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater that 800 watts,
then forced convection must be used.

Warning Install the instrument according to the enclosure protection provided.
This instrument does not protect against the ingress of water. This
iustrument protects agaius finger access to hazardous parts within the
enclosure.




Compliance with German FTZ Emissions Requirements
This network analyzer complies with German FlZ 526/527 Radiated Emissions and Conducted
Emission requirements.




Compliance with German Noise Requirements
This is to declare that this instrument is in conformance with the German Regulation on
Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordung -3. GSGV
Deutschland).

Acoustic Noise Emission/Geraeuschemission
LpA<70 dB Lpa<70 dD
User's Guide Overview
n Chapter 1, "HP 8753E Description and Options, n describes features, functions, and available
options.
n Chapter 2, "Making Measurements," contains step-by-step procedures for making
measurements or using particular functions.
n Chapter 3, "Making Mixer Measurements, n contains step-by-step procedures for making
calibrated and error-corrected mixer measurements.
w Chapter 4, "Printing, Plotting, and Saving Measurement Results," contains instructions
for saving to disk or the analyzer internal memory, and printing and plotting displayed
measurements.
n Chapter 5, "Optimizing Measurement Results, n describes techniques and functions for
achieving the best measurement results
n Chapter 6, "Application and Operation Concepts, n contains explanatory-style information
about many applications and analyzer operation.
n Chapter 7, "Specifications and Measurement Uncertainties," defines the performance
capabilities of the analyzer.
n Chapter 8, "Menu Maps," shows softkey menu relationships.
n Chapter 9, "Key Dell&ions," describes all the front panel keys, softkeys, and their
corresponding HP-IB commands.
n Chapter 10, "Error Messages," provides information for interpreting error messages
n Chapter 11, "Compatible Peripherals, n lists measurement and system accessories, and
other applicable equipment compatible with the analyzer. Procedures for configuring the
peripherals, and an HP-IB programming overview are also included.
n Chapter 12, "Preset State and Memory Allocation," contains a discussion of memory
allocation, memory storage, instrument state definitions, and preset conditions.
n Appendix A, "The CITIlile Data Format and Key Word Reference, n contains information on
the CITIGle data format as well as a list of CITIflle keywords
n Appendix B, "Determining System Measurement Uncertainties," contains information on how
to determine system measurement uncertainties.




lx
Network Analyzer Documentation Set
The Installation and Quick Start Guide
familiarizes you with the network analyzer's
front and rear panels, electrical and
environmental operating requirements, as well
as procedures for installing, configuring, and
verifying the operation of the analyzer.




The User's Guide shows how to make
measurements, explains commonly-used
features, and tells you how to get the most
performance from your analyzer.




The Quick Reference Guide provides a
summary of selected user features.




The HEW3 Programming and Command
Reference Guide provides programming
information for operation of the network
analyzer under HP-IB control.




X
The HP BASIC Programming Examples
Guide provides a tutorial introduction using
BASIC programming examples to demonstrate
the remote operation of the network analyzer.




The System Vertication and `lkst Guide
provides the system verification and
performance tests and the Performance Test
Record for your analyzer.




xl
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and EN 45014


Manufacturer's Name:
Hewlett-Packard Co. Hewlett-Packard Japan, Ltd.

Manufacturer's Address:
Microwave Instruments Division Kobe Instrument Division
1400 Fountaingrove Parkway l-3-2, Murotani, Nishi-ku, Kobe-shi
Santa Rosa, CA 95403-I 799 Hyogo, 651-22
USA Japan


Declares that the product:

Product Name: Network Analyzer

Model Number: HP 8753E

Product Options: This declaration covers all options of the above
product

Conforms to the following Product specifications:

Safety: IEC 61010-1:199O/EN 61010-I:1993
CAN/CSA-C22.2 No. 1010.1-92

EMC: CISPR 11:1990/EN 5501 I:1991 Group I, Class A
IEC 801-2:199l/EN 50082-I:1992 4 kV CD, 8 kV AD
IEC 801-3:1984/EN 50082-I:1992 3 V/m, 27-500 MHz
IEC 801-4:1988/EN 50982-I :I 992 0.5 kV sig. lines, 1 kV power lines

Supplenientary Information:

The product herewith complies with the requirements of the Low Voltage Directive
73/23/EEC and the EMC Directive 89/336/EEC and carries the CE-marking accordingly.



c


&A


Jo HiaWQuality Engineering Manager Mike Obara/Quality Engineering Manager
d
Santa Rosa, 21 Jan. 1998 Kobe, 14 Jan. 1998


Europeen Contact: Your knxl Hewlett-Packard Sales and Servke Oftice or Hewlett-Packard GmbH Department HQ-
TFIE, Heneneberger Strasse 130. D71034 Boblingen, Germany (FAX +49-7031-U-3143




xii
Contents
1. HP 8753E Description and Options
Where to Look for More Information . . . . . . . . . . . . . . . . . . . . . l-l
Analyzer Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-2
Front Panel Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-4
Analyzer Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-7
Rear Panel Features and Connectors . . . . . . . . . . . . . . . . . . . . . l-11
Analyzer Options Available . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Option lD5, High Stability Frequency Reference . . . . . . . . . . . . . . . 1-13
Option 002, Harmonic Mode . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Option 006,6 GHz Operation . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Option 010, Time Domain . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Option 011, Receiver Configuration . . . . . . . . . . . . . . . . . . . . . 1-13
Option 075,750 Impedance . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Option lDT, Delete Display . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Option lCM, Rack Mount Flange Kit Without Handles . . . . . . . . . . . . 1-13
Option lCP, Rack Mount Flange Kit With Handles . . . . . . . . . . . . . . 1-14
Service and Support Options . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
Differences among the HP 8753 Network Analyzers . . . . . . . . . . . . . . 1-15

2. B