Text preview for : M9370-90002 M937xA PXIe Vector Network Analyzer Modules - Data Sheet and Technical Specs c20140923 [ part of



| Home

Keysight
PXIe Vector Network
Analyzers
M9370A 300 kHz to 4 GHz
M9371A 300 kHz to 6.5 GHz
M9372A 300 kHz to 9 GHz
M9373A 300 kHz to 14 GHz
M9374A 300 kHz to 20 GHz
M9375A 300 kHz to 26.5 GHz




Data Sheet and
Technical
Specifications
Electronic calibration (ECal): enables fast,
Drive down the size of test easy and accurate calibrations
When you need to measure basic S-parameters,
the right mix of speed, performance and footprint Maintain hardware compliance
gives you an edge. Sharpen your edge with the
Keysight PXI vector network analyzer (VNA), ideal The M937xA is PXI compliant and designed to
for: benefit from fast data interfaces. The PXI VNA can
be integrated with other test and automation
Wireless manufacturing test modules in either a PXIe or Hybrid slot. The PXI
Handset component test format offers high performance in a small, rugged
Aerospace/defense depot maintenance package. It is an ideal deployment platform for
level testing many automated test systems.
Aerospace/defense manufacturing test

Get more VNA in less space
The Keysight PXI VNA is a full two-port VNA that
fits in just one slot. The PXI VNA also performs
fast, accurate measurements and reduces your
cost-of-test by letting you simultaneously
characterize many devices -- two-port or
multiport -- using a single PXI chassis.
It offers the best PXI VNA performance on key
specifications such as dynamic range,
measurement speed, and trace noise. Each
module is a completely independent two-port
network analyzer and up to 16 modules can be
added to a chassis for multi-site and multiport
applications. All ports are fully synchronous so
multiple ports can be measured simultaneously
with multiport error correction applied. For more
information, see page 28.

Achieve confidence and continuity in
measurements
The M937xA extends Keysight's expertise in
metrology to the modular form factor. It provides
the same quality results you have come to expect
in our vector network analyzers. Here are a few
examples of ways that we continue to provide
measurement quality and continuity.
Dynamic range: > 114 dB at 9 GHz,
> 110 dB at 20 GHz
Trace noise: < 0.003 dB
Sweep speed: 18 to 33 msec across 401
points Pay for only the frequency range you need with
Maximum leveled power: up to +7 dBm the widest choice of frequency ranges from 300
Stability: 0.005 dB/