Text preview for : 4225366C_w.pdf part of Fluke 4225366C w Fluke 5502A 4225366C_w.pdf



Back to : 4225366C_w.pdf | Home

5502A Multi-Product
Calibrator
Extended specifications
General Specifications
The following tables list the 5502A specifications. All specifications are valid after allowing a warm-up period of
30 minutes, or twice the time the 5502A has been turned off. (For example, if the 5502A has been turned off for
5 minutes, the warm-up period is 10 minutes.)
All specifications apply for the temperature and time period indicated. For temperatures outside of tcal 5 C (tcal is the
ambient temperature when the 5502A was calibrated), the temperature coefficient as stated in the General Specifications
must be applied.
The specifications also assume the Calibrator is zeroed every seven days or whenever the ambient temperature changes
more than 5 C. The tightest ohms specifications are maintained with a zero cal every 12 hours within 1 C of use.
Also see additional specifications later in this chapter for information on extended specifications for ac voltage and
current.
Warmup Time .....................................................Twice the time since last warmed up, to a maximum of 30 minutes.
Settling Time ......................................................Less than 5 seconds for all functions and ranges except as noted.
Standard Interfaces ............................................IEEE-488 (GPIB), RS-232
Temperature
Operating .........................................................0 C to 50 C
Calibration (tcal) ..............................................15 C to 35 C
Storage ............................................................-20 C to +70 C; The DC current ranges 0 to 1.09999 A and 1.1 A
to 2.99999 A are sensitive to storage temperatures above 50 C. If
the 5502A is stored above 50 C for greater than 30 minutes,
these ranges must be re-calibrated. Otherwise, the 90 day and 1
year uncertainties of these ranges double.
Temperature Coefficient ....................................Temperature coefficient for temperatures outside of tcal