Text preview for : 595_903_01A.pdf part of Keithley 595 903 01A Keithley 595 595_903_01A.pdf



Back to : 595_903_01A.pdf | Home

Model 595 Quasistatic CV Meter
Quick Reference Guide




A GREATER MEASURE OF CONFIDENCE
INTRODUCTION

This reference and programmingguide contains condensed
specifications, descriptions of the various features and in-
formation for using the Model 595. Also, IEEE-488program-
ming information and several example programs written for
commonly used controllers are also included to get the
Model 595 "up and running" on the bus.


The Model 595 QuasistaticCV Meter measures capacitance
using a unique feedback charge method. This method cor-
rects capacitance readings for background leakage current.
A built-in IEEE-488 interface, plotter driver and labeled grid
program enable the Model 595 to plot IV or quasistatic CV
curves for a variety of semiconductor materials and
components.




01986, Keithley Instruments, Inc.
Instrument Division
Cleveland, Ohio, U.S.A.
Document Number: 595-903-01 Rev. A
1
TABLE OF CONTENTS

CONDENSED SPECIFICATIONS .................. 3
MODEL 595 FEATURES ..................... 7-20
FRONT PANEL PROGRAMS ................. 21-26
MEASUREMENT CONNECTIONS............. 29-34
IEEE PROGRAMMING...................... 35-49
PROGRAMS ................................ 50
IBM PC or XT . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Applell ............................... 53
HP85 ................................ 55
HP9816 .............................. 57
HP9825A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
DEC E 11 ............................
l 61
PETICBM 2001 ......................... 63




2
CONDENSED SPECIFICATIONS

CAPACITANCE

Maximum Allowable
Accuracy' W at Half Range C
t
Reso- (1 b a r ) 18-28OC Delay Time=O.Oirsec
Range lution +(%rdg+counts) Step V=O.lV

200pF lOfF 1.0 + 10 9O.OOOpA
2nF lOOfF 0.8 + 2 0.900nA
20nF OF 0.6 + 2 9.000nA

*Exclusive of noise, for Step V 2 0.05V and
Delay Time 5 one second.


CURRENT

Temperature
Accuracy' Coefficient
Reso- (1 Bar) 18-28OC 0-18OC & 28-4OoC
Range lution ? (%rdg+counts) f (%rdg+ counts)l