Text preview for : CMD60 1050.9008.60 Chapter 1.pdf part of Rohde & Schwarz CMD60 1050.9008.60 Chapter 1 Rohde & Schwarz R&S CMD60 SERIES DOCS CMD60 1050.9008.60 Chapter 1.pdf



Back to : CMD60 1050.9008.60 Chapte | Home

Test and
Measurement Division




Operating Manual



Digital Radiocommunication
Tester for DECT
CMD60
1050.9008.60




Printed in the Federal
Republic of Germany




1051.5929.12-05- 1
CMD60 Tabbed Divider Overview




Data Sheet

Safety Instructions
Certificate of quality
EC Certificate of Conformity
List of R & S Representatives




1 Preparation for Use ............................................................................. Tabbed Divider 1

2 Manual Operation ................................................................................. Tabbed Divider 2
Getting Started
Description of the Menus (Reference Description)


3 Remote Control ..................................................................................... Tabbed Divider 3

4 Maintenance ........................................................................................... Tabbed Divider 4

5 Testing the Rated Specifications................................................... Tabbed Divider 5

Annex A (Interfaces) .............................................................................................. Tabbed Divider 6

Annex B (List of Error Messages) ................................................................... Tabbed Divider 7

Annex C (List of Commands) ............................................................................ Tabbed Divider 8

Annex D (Program Example) ............................................................................. Tabbed Divider 9

Index .......................................................................................................................... Tabbed Divider 10




1051.5929.12 3 E-4
Safety Instructions
This unit has been designed and tested in accordance with the EC Certificate of Conformity and has left the
manufacturer's plant in a condition fully complying with safety standards.
To maintain this condition and to ensure safe operation, the user must observe all instructions and warnings
given in this operating manual.
1. The unit may be used only in the operating 7. It is not permissible to interrupt the PE conduc-
conditions and positions specified by the tor intentionally, neither in the incoming cable
manufacturer. Unless otherwise agreed, the fol- nor on the unit itself as this may cause the unit
lowing applies to R&S products: to become electrically hazardous.
Pollution severity 2, overvoltage category 2, IP Any extension lines or multiple socket outlets
degree of protection 2X, altitude max. 2000 m. used must be checked for compliance with rele-
The unit may be operated only from supply net- vant safety standards at regular intervals.
works fused with max. 16 A. 8. If the unit has no power switch for disconnection
2. For measurements in circuits with voltages Vrms from the AC supply, the plug of the connecting
> 30 V, suitable measures should be taken to cable is regarded as the disconnecting device.
avoid any hazards. In such cases it must be ensured that the power
(using, for example, appropriate measuring plug is easily reachable and accessible at all
equipment, fusing, current limiting, electrical times (length of connecting cable approx. 2 m).
separation, insulation). Functional or electronic switches are not suit-
able for providing disconnection from the AC
3. If the unit is to be permanently wired, the PE supply.
terminal of the unit must first be connected to If units without power switches are integrated in
the PE conductor on site before any other con- racks or systems, a disconnecting device must
nections are made (installation and cabling of be provided at system level.
the unit to be performed only by qualified techni-
cal personnel). 9. Applicable local or national safety regulations
and rules for the prevention of accidents must
4. For permanently installed units without built-in be observed in all work performed.
fuses, circuit breakers or similar protective de-
Prior to performing any work on the unit or
vices, the supply circuit must be fused such as
opening the unit, the latter must be discon-
to provide suitable protection for the users and
nected from the supply network.
equipment.
Any adjustments, replacements of parts, main-
5. Prior to switching on the unit, it must be ensured tenance or repair may be carried out only by
that the nominal voltage set on the unit matches authorized R&S technical personnel.
the nominal voltage of the AC supply network. Only original parts may be used for replacing
If a different voltage is to be set, the power fuse parts relevant to safety (eg power switches,
of the unit may have to be changed accordingly. power transformers, fuses). A safety test must
6. Units of protection class I with disconnectible be performed after each replacement of parts
AC supply cable and appliance connector may relevant to safety.
be operated only from a power socket with (visual inspection, PE conductor test, insulation-
earthing contact and with the PE conductor con- resistance, leakage-current measurement, func-
nected. tional test).
10. Any additional safety instructions given in this
manual are also to be observed.

Safety-related symbols used on equipment and documentation from R&S:




Observe Weight PE terminal Ground Danger! Warning! Ground
operating indication for terminal Shock hazard High
instructions units >18 kg temperatures
Warning!
Hot surfaces




095.1000 Sheet 17
EC Certificate of Conformity




Certificate No.: 9502035


This is to certify that:


Equipment type Order No. Designation


CMD60 1050.9008.60 Dig. Radiocommunication Tester


complies with the provisions of the Directive of the Council of the European Union on the
approximation of the laws of the Member States

- relating to electrical equipment for use within defined voltage limits
(73/23/EEC revised by 93/68/EEC)

- relating to electromagnetic compatibility
(89/336/EEC revised by 91/263/EEC, 92/31/EEC, 93/68/EEC)

Conformity is proven by compliance with the following standards:

EN61010-1 : 1993 + A2 : 1995
EN50081-1 : 1992
EN50082-1 : 1992

Affixing the EC conformity mark as from 1995



ROHDE & SCHWARZ GmbH & Co. KG
M