*TB 9-6625-1957-35
SUPERSEDED COPY DATED 18 JULY 1988
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
DIGITAL MULTIMETER, TEKTRONIX TYPES
DM 501 AND DM 501A
WITH DC HIGH VOLTAGE PROBE, BALLANTINE,
MODEL 10800C; JOHN FLUKE, MODEL 80K-40;AND
TEKTRONIX, TYPE 010-0277-00
Headquarters, Department of the Army, Washington, DC
22 February 1993
Approved for public release; distribution is unlimited
REPORTING OF ERRORS
You can help improve this publication by calling attention to errors and by
recommending improvements and stating your reasons for the
recommendations. Your letter or DA Form 2028, Recommended Changes to
Publications, should be mailed directly to Commander, U.S. Army Aviation and
Missile Command, ATTN: AMSAM-TMD-EP, Redstone Arsenal, AL 35898-5000.
FAX to DSN 788-2313 (commercial 205-842-2313). A reply will be furnished
directly to you.

Paragraph Page
SECTION I. IDENTIFICATION AND DESCRIPTION
Test instrument identification ............................ 1 2
Forms, records, and reports ................................ 2 2
Calibration description........................................ 3 3
SECTION II. EQUIPMENT REQUIREMENTS
Equipment required ............................................ 4 4
Accessories required............................................ 5 4
III. CALIBRATION PROCESS FOR DIGITAL
MULTIMETER, TEKTRONIX, TYPE DM 501
Preliminary instructions ..................................... 6 5
Equipment setup ................................................. 7 6
Dc voltage ............................................................ 8 6
Resistance............................................................ 9 7
Dc current............................................................ 10 8
Ac voltage ............................................................ 11 8
Power supply ....................................................... 12 9
Final procedure ................................................... 13 10



__________
*This bulletin supersedes TB 9-6625-1957-35, dated 18 July 1988, including all changes.
TB 9-6625-1957-35

Paragraph Page
SECTION IV. CALIBRATION PROCESS FOR DIGITAL
MULTIMETER, TEKTRONIX, TYPE DM 501A
Preliminary instructions ................................... 14 10
Equipment setup 15 10
Dc voltage .......................................................... 16 11
Ac voltage .......................................................... 17 12
Ac dBm/dBV ...................................................... 18 13
Dc current.......................................................... 19 14
Resistance.......................................................... 20 14
Final procedure ................................................. 21 15
V CALIBRATION PROCESS FOR DC HIGH
VOLTAGE PROBE, BALLANTINE,
MODEL 10800C, JOHN FLUKE, MODEL
80K-40, AND TEKTRONIX, TYPE
010-0277-00
Preliminary instructions ..................................... 22 15
Equipment setup ............................................... 23 16
Dc probe check................................................... 24 16
Final procedure ................................................. 25 17


SECTION I
IDENTIFICATION AND DESCRIPTION

1. Test Instrument Identification. This bulletin provides instructions for the
calibration of Digital Multimeter, Tektronix, Types DM 501 and DM 501A with DC High
Voltage Probe, Ballantine, Model 10800C; John Fluke, Model 80K-40; and Tektronix, Type
010-0277-00. The manufacturer's manuals were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.

a. Model Variations. The John Fluke, Model 80K-40 is the same as the Tektronix,
Type 010-0277-00 probe and is nonadjustable.

b. Time and Technique. The time required for this calibration is approximately 1.5
hours, using the dc and low frequency technique.

2. Forms, Records, and Reports

a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.

b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).


2
TB 9-6625-1957-35
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.

Table 1. Calibration Description
Test instrument parameters Performance specifications
Tektronix, Type DM 501
Dc voltage Range: 0 to 1000 V (in 4 ranges)
Accuracy: