1265775G_w.pdf | | Oscilloscope Calibration Options for
55XX Series
Multi-Product Calibrators
Extended
Specifications
These specifications apply
to the 5520A-SC1100, 5500A-
SC600 and 5500A-SC300
Oscilloscope Calibration Options.
For general specifications, see
the Extended Specifications for
the individual calibrator of interest
(downloadable from the Fluke
Calibration web site or available
from your Fluke Calibration
representative).The specifications
are valid when the calibrator is
operated under the conditions
specified in the operator manual,
and has completed a warm-up
period of at least twice the length
of time the calibrator was pow-
ered off, up to a maximum of 30
minutes.
The 5520A-SC1100, with 1100
MHz bandwidth, can be installed
in a 5522A calibrator. The 5500A-
SC600, with 600 MHz bandwidth,
can be installed in any 55XX Series
Multi-Product Calibrator. The 300
MHz option, the 5500A-SC300, is
only c |