| File information: | |
| File name: | GFG-813_ADJ[1].pdf [preview GFG-813 ADJ[1]] |
| Size: | 52 kB |
| Extension: | |
| Mfg: | GW-INSTEK |
| Model: | GFG-813 ADJ[1] 🔎 |
| Original: | GFG-813 ADJ[1] 🔎 |
| Descr: | GW-INSTEK GFG-813_ADJ[1].pdf |
| Group: | Electronics > Other |
| Uploaded: | 28-02-2020 |
| User: | Anonymous |
| Multipart: | No multipart |
| Information about the files in archive: | ||
| Decompress result: | OK | |
| Extracted files: | 1 | |
File name GFG-813_ADJ[1].pdf PE-C4-410 PAGE 1 GOODWILL INSTRUMENT ( MALAYSIA ) SDN. BHD. CALIBRATION STANDARD PROCEDURES FUNCTION GENERATOR MODEL : GFG-813 APPLICABLE ODM MODEL : FG-813; BK-3040 GW GOODWILL INSTRUMENT ( M ) SDN BHD. APPROVED BY CHECKED BY PREPARED BY TOTAL PAGE : 15 ( INCLUDING COVER PAGE ) FOR INTERNAL USE ONLY PAGE : 2 CALIBRATION PROCESS MODEL: GFG-813 STATION : PRE-ADJUST NO Adjustment Location Particular Specification Remarks 1 Initial Setting. POWER Switch = ON; Instrument Used : EXT COUNTER : release all buttons; 20Mhz O'scope MAIN GENERATOR : AMPL=Max, ATT= Release, OFFSET & SYM= Push, TRIGGER PHASE=Free run; MODULATION : AM, FM & SWP=Release, MODULATION Output=Max, SYM=CAL. 2 Puncture [ HI POT] Test. Connect AC cord's GND, Life & Neutral pins respective. GPT-515A : O/P voltage = 1.5 KV Cutoff Current = 1 mA Timer = 60 Sec 3 Functioning Check. Check the function of all VRs & button switches 20Mhz O'scope corresponing to the output waveform. 4 Burn - in. Deliver the sets to the burn-in chamber. 50 | ||

| Date | User | Rating | Comment |