| File information: | |
| File name: | 5725A___mveng0000.pdf [preview 5725A___mveng0000] |
| Size: | 16 kB |
| Extension: | |
| Mfg: | Fluke |
| Model: | 5725A___mveng0000 🔎 |
| Original: | 5725A___mveng0000 🔎 |
| Descr: | Fluke 5700a 5725A___mveng0000.pdf |
| Group: | Electronics > Other |
| Uploaded: | 04-09-2019 |
| User: | Anonymous |
| Multipart: | No multipart |
| Information about the files in archive: | ||
| Decompress result: | OK | |
| Extracted files: | 1 | |
File name 5725A___mveng0000.pdf Instrument Security Procedures Model: Fluke 5725A Product Name: Amplifier Instrument Description: The 5725A amplifier is used to increase the current and voltage capability for the Fluke 5700A/5720A and Fluke 5500A. Memory Description: The 5725A amplifier contains volatile and non-volatile memory. The 5725A amplifier has 2K x 8 Static Ram, 2K x 8 EEPROM and a factory-programmed ROM. The EEPROM is used to store calibration constants. This memory is not user accessible or configurable. Memory Cleaning Instructions: Static RAM The Static Ram is volatile memory and is erased when the instrument is no longer powered. EEPROM The EEPROM can be set to factory default calibration constants using a Fluke 5700A/5720A. Process using Fluke 5700A/5720A: 1. Set the CALIBRATION switch on the back of the 5725A to ENABLE. 2. From the front panel of the 5720A/5720A, select "Setup Menus" softkey. 3. Select "Instrmt Setup" softkey. 4. Select "Format EEPROM" softkey. 5. Select 5725 softkey. 6. Select "YES" softkey. 5725A SOMV 1 of 1 10/13/2009 | ||

| Date | User | Rating | Comment |