Now downloading free:Fluke 4225366C w

Fluke 4225366C w free download

Various electronics service manuals

File information:
File name:4225366C_w.pdf
[preview 4225366C w]
Size:3164 kB
Extension:pdf
Mfg:Fluke
Model:4225366C w 🔎
Original:4225366C w 🔎
Descr: Fluke 5502A 4225366C_w.pdf
Group:Electronics > Other
Uploaded:11-12-2019
User:Anonymous
Multipart:No multipart

Information about the files in archive:
Decompress result:OK
Extracted files:1
File name 4225366C_w.pdf

5502A Multi-Product Calibrator Extended specifications General Specifications The following tables list the 5502A specifications. All specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5502A has been turned off. (For example, if the 5502A has been turned off for 5 minutes, the warm-up period is 10 minutes.) All specifications apply for the temperature and time period indicated. For temperatures outside of tcal 5 C (tcal is the ambient temperature when the 5502A was calibrated), the temperature coefficient as stated in the General Specifications must be applied. The specifications also assume the Calibrator is zeroed every seven days or whenever the ambient temperature changes more than 5 C. The tightest ohms specifications are maintained with a zero cal every 12 hours within 1 C of use. Also see additional specifications later in this chapter for information on extended specifications for ac voltage and current. Warmup Time ...Twice the time since last warmed up, to a maximum of 30 minutes. Settling Time ...Less than 5 seconds for all functions and ranges except as noted. Standard Interfaces ...IEEE-488 (GPIB), RS-232 Temperature Operating ...0 C to 50 C Calibration (tcal) ...15 C to 35 C Storage ...-20 C to +70 C; The DC current ranges 0 to 1.09999 A and 1.1 A to 2.99999 A are sensitive to storage temperatures above 50 C. If the 5502A is stored above 50 C for greater than 30 minutes, these ranges must be re-calibrated. Otherwise, the 90 day and 1 year uncertainties of these ranges double. Temperature Coefficient ...Temperature coefficient for temperatures outside of tcal

>> View document online <<



>> Download document << eServiceInfo Context Help



Was this file useful ? Share Your thoughts with the other users.

User ratings and reviews for this file:

DateUserRatingComment

Average rating for this file: 0.00 ( from 0 votes)


Similar Service Manuals :
Fluke lt1097 - Fluke Initial F5700A Cal Dump - Fluke DN2540 - Fluke 5790a mveng0000 - Fluke at25010an10sq27 - Fluke Readme - Fluke 74HC HCT4052 -
 FB -  Links -  Info / Contacts -  Forum -   Last SM download : TCL & SINGER 21N26NF

script execution: 0.03 s